Operation
14
Instructions for Use ERGOeco
BOWA-IFU-13120-ERGOECO-S0-EN-20220420
7.3 During the surgery
Introducing the instrument
1. Close the handle in order to keep the jaws closed.
2. Insert the instrument into the trocar sleeve.
Grasping, clamping and sealing tissue
1. Position the jaws at the surgical site.
2. Turn the rotary wheel (with the jaws open) to align the jaws to the target tissue.
3. Place the tissue to be coagulated between the jaws.
4. Close the jaws to grasp the tissue. Never grasp too much tissue.
5. Use the foot switch to activate the HF current of the HF generator for coagulation:
A continuous acoustic signal is emitted during the entire activation period.
WARNING
Risk of patient injury due to tissue pinching, especially with limited visibility!
Expose the tissue to be coagulated as much as possible to prevent it from
becoming pinched inadvertently!
Only perform the operation if you have clear visibility.
WARNING
Risk of patient injuries from hot jaws and release of steam!
Instrument tips may still be hot immediately after HF power has been
switched off.
Maintain sufficient distance between instrument tips and sensitive
tissue structures, such as nerves, pancreas or intestines.
Do not use hot laparoscopy instruments to prepare the tissue.
Do not place any laparoscopy instrument on the patient.
WARNING
Risk of patient injury due to accidental activation of the laparoscopy instrument!
Never use the AUTOSTART function.
Do not switch on the HF current before the jaws are in contact with
the tissue to be coagulated.
Activating the instrument inadvertently can cause injuries to the patient.