Sensors Configuration
For each sensor to be connected to its dedicated input channel, it is
possible to set the following parameters:
•
“
Identificativo
” / ID
Enter an alpha-numerical name (up to 12
characters) to assign an ID to identify the type of sensor:
Click on Identificativo
(ID) of
“
CANALE
0”
:
the field will become
yellow to indicate that
it is now active. Then
enter the new name
and confirm by
pressing
“Invio”
Figure 10
–
Configuring the unit
•
“
Unità
”
this defines the engineering unit to be associated to the
type of reading / sensor being used (e.g. mm, °C, KN, mA)
Proceed as above for the remaining sensors / channels
Note
Channels 1 to 12 (i.e.
“
Canale
0”
to
“
Canale
11”
), being part of the same type of
measurement, will all have the same ID and Engineering unit /
“Unit
à
”
.
Therefore, the parameters for channels
’
sensors 1 to 11 will be imported from the
corresponding values as entered at
“
Canale
0”
•
“
Precisione
”
Defines the number of decimals of your interest
Proceed as above for the remaining sensors / channels
Note
The decimals to be set can be selected from a drop-down list. Choose the decimals
according to the precision of the sensor being connected.
•
“
Calibrazione
”
This field allows entering up to 10 calibration
points for the best conversion of electric values from the sensors
into engineering units displayed by the software.