Wash, but do not submerge, the pulse generator and leads to remove
body
fl
uids and debris using a disinfectant solution. Do not allow
fl
uids to
enter the pulse generator’s lead ports.
•
Use a Boston Scienti
fi
c Returned Product Kit to properly package the
pulse generator and send it to Boston Scienti
fi
c.
SPECIFICATIONS
Speci
fi
cations (Nominal)
Table 5. Model Number and Lead Length
Model
Single Coil/Dual
Coil
ePTFE-covered
coil(s)
Length
0675
Dual Coil
No
59 cm
0676
Dual Coil
No
64 cm
0692
Single Coil
Yes
59 cm
0693
Single Coil
Yes
64 cm
0657
Single Coil
Yes
70 cm
0695
Dual Coil
Yes
59 cm
0696
Dual Coil
Yes
64 cm
0658
Dual Coil
Yes
70 cm
Table 6. Speci
fi
cations (Nominal)
Characteristic
Nominal
Terminal type
DF4-LLHH (dual-coil models)
DF4-LLHO (single-coil models)
Compatibility
Pulse generators with a DF4-LLHH
port or GDT-LLHH port, which
accepts either a DF4-LLHH or
DF4-LLHO terminal
Fixation
Extendable/retractable helix
Expected number of rotations to fully
extend/retract the helix
a
11 turns
Maximum number of rotations allowed to
extend/retract the helix
20 turns
Maximum
fi
xation helix penetration depth
1.9 mm
Electrode:
Distal coil surface area
450 mm
2
31
Outdated version. Do not use.
Version überholt. Nicht verwenden.
Version obsolète. Ne pas utiliser.
Versión obsoleta. No utilizar.
Versione
obsoleta. Non utilizzare.
Verouderde
versie. Niet gebruiken.
Föråldrad
version. Använd ej.
Παλιά
έκδοση
.
Μην
την
χρησιμοποιείτε
.
Versão obsoleta. Não utilize.
Forældet version. Må ikke anvendes.
Zastaralá verze. Nepoužívat.
Utdatert versjon. Skal ikke brukes.
Zastaraná verzia. Nepoužíva
ť
.
Elavult verzió. Ne használja!
Wersja nieaktualna.
Nie u
ż
ywa
ć
.