Pump is under control of the Maestro 4000
™
. This
mode is entered when the Maestro 4000 and pump are
powered on and communicating; and an open-irrigated
(OI) catheter is connected.
•
No action required.
Maestro 4000 has a diagnostic error, or pump lost com-
munication with Maestro during RF delivery.
•
Check the Maestro 4000 for a diagnostic error and
resolve the error according to the Maestro 4000
Operator manual. If there is no error and Maestro
4000 is delivering RF energy when this message
is displayed, check connection of the communica-
tion cable between the MetriQ and Maestro 4000.
Tighten thumbscrews on connectors.
After the power on self-test is complete, the Pump
will prompt the user to select “Start New Case” or
“Continue Previous Case”
•
Select “Continue Previous Case” to keep the previ-
ous dispensed and infused volume levels.
Note: If
the saline bag size is smaller than 2000 mL you
must also reset the bag size through the menu.
Otherwise the remaining volume will be incorrect.
The pump door is open and a Standby, Low, High or
Purge flow button was pressed.
(This message has the
same text as P04, but is only displayed when there is
no flow).
•
Close the door, then press the desired flow button.
If door is already closed it may not be completely
latched. Twist the door knob to the right.
Pump lost communication with the Maestro 4000. This
could occur if the communications cable is not securely
attached to both devices. It will also occur if the
Maestro 4000 has a Safe state error. This message is
only displayed if the pump is in Automatic Mode when
the error occurs.
•
Check the Maestro 4000 for errors. Refer to the
Maestro 4000 manual to resolve generator errors.
•
If message is displayed while the communications
cable is connected, do the following:
1. Check the cable connections between the
Maestro 4000 and MetriQ. Tighten thumb screws
on the connectors.
2. Replace the communications cable.
Displayed if the Pump is not under control of a Maestro
4000. The pump will also be in manual mode if the
pump and Maestro 4000 are connected, but an OI cath-
eter is not attached to the Maestro 4000.
•
If you intend to operate the pump in automatic
mode, but the pump shows “Manual Mode”,
first check to make sure the OI catheter and pod
are securely connected. Next check the cable as
described under “
GENERATOR DISCONNECTED
”
The fluid remaining in the saline bag is below the menu
setting for “Low Fluid Warning”.
•
Check bag to verify saline volume. User error may
cause a discrepancy between actual and calcu-
lated. If low, press the stop button to stop flow.
Change saline bag. Then go to menu and set the
parameter “New Saline Bag” to Yes.
After the power on self-test, the Pump will prompt the
user to select “Start New Case” or “Continue Previous
Case”
•
Select “Start New Case” to reset the dispensed
and infused volume levels to zero.
An unrecoverable error has occurred on the pump.
Pump enters
SAFE
mode and ignores all input.
•
Record the error number, power off the pump and
contact BSC field service for further assistance.
Tubing is not loaded correctly.
•
Open the pump door. Check that the tubing “stops”
are in the recesses on either side of the pump
head.
Automatic Mode
Manual Mode
SALINE VOL. LOW < XXX ML
Start New Case
SYSTEM FAULT
TUBING NOT LOADED
CHECK GENERATOR
Continue Previous Case
COVER OPEN
GENERATOR DISCONNECTED
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Cause:
Action:
Operational Messages
Operational messages explain MetriQ
™
Pump behavior or indicate a non-hazardous condition. Not all operational messages are listed here.
For a description of all messages see the Operator’s Manual.
All cited trademarks are the property of their respective owners. CAUTION: The law restricts these devices to sales by or on the
order of a physician. Indications, contraindications, warnings and instructions for use can be found in the product labeling
supplied with each device. Information for the use only in countries with applicable health authority product registrations.
EP-252701-AA JUN2014 Printed in Germany by medicalvision.
www.bostonscientific.com
© 2014 Boston Scientific Corporation
or its affiliates. All rights reserved.
DINEP2212EA
44
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
43