![Boston Scientific Maestro 3000 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/boston-scientific/maestro-3000/maestro-3000_connection-manual_2796091016.webp)
IntellaTip MiFi
™
XP Ablation Catheter
Advanced Set Up with Ensite™ Velocity™ Cardiac Mapping System
Step 1. Connecting the IntellaTip MiFi XP Conventional Bipoles as a standard ablation catheter
a. Connect Dispersive Surface Electrode(s) to the Velocity EPT GenConnect Module.
b. Connect M004
620
0 Cable to the Filter Module (yellow plug) and the Velocity EPT GenConnect (red plug).
c. Connect the M004
620
0 Cable
(black plug)
into the IntellaTip MiFi XP Catheter.
d. Connect Dispersive Interconnect Cable (SJM 100010915) to the Maestro Pod Dispersive Pad Inputs.
e. Connect Catheter Interconnect Cable (SJM 100010397) to the Maestro Pod Catheter Input.
f. Connect the M004
653S
0 Cable into the Pod. Pin out conventional Bipoles into the Pin Connection Box
Step 2. Connecting the IntellaTip MiFi XP Mini Electrodes as a separate diagnostic catheter
a. Connect the M004
653S
0 Cable plug into the green port marked “Recording System” on the Filter Module.
Connect the pins to the Catheter Pin Connection Box.
b. Pin out the Mini Electrodes (MEs) - D, 2 and 3. Pin 4 is not used, but can be plugged into any unused channel.
c. Do Not Connect the Reference Cable (M004 3636 0) when using the EnSite Velocity System.
Step 3. Defining the IntellaTip MiFi XP Catheter Parameters
a. The IntellaTip MiFi Catheter should be defined as both an ablation (ABL) and diagnostic (MIFI) catheter in the
Catheter Set Up Screen.
b. Input waveforms: D-2, 2-3 and 3-D, for the IntellaTip MiFi XP Catheter Mini Electrodes.
(Green)
(Black)
(Yellow)
Velocity EPT GenConnect
Catheter Pin Connection Box
Pannel on the front
of the Amplifier
IntellaTip MiFi Filter Module
M004 1212 0
IntellaTip MiFi XP Catheter
Maestro 3000 Cardiac Ablation Controller
RF Ablation Pod
M004 21860T 0
Dispersive Indifferent Patch
(Commercially Available)
IntellaTip MiFi XP Cable
M004 620 0
M004 653S 0
Cable: Filter Module
to Recorder
(Red)
Cable: Pod to Recorder
M004 653S 0
NAME
ABL
MIFI
Electrodes
4
3
Type:
Ablation
Diagnostic
Diameter:
8 Fr
(2.66 mm)
8 Fr
(2.66 mm)
Distal Length:
8/10 mm
1 mm
Electrode Length:
1.27 mm
1 mm
Electrode Spacing:
2.5-2.5-2.5
1-1
16
IntellaTip MiFi
™
XP
Temperature Ablation Catheter
System Set Up with Stockert EP-Shuttle RF Generator
IntellaTip MiFi
™
XP
Ablation Catheter
Advanced Set Up with Ensite™ Velocity™ Cardiac Mapping System
15