19
Charging batteries
If the device is not used for an extended period (4 weeks or more), set
the switch in the internal battery compartment to OFF and remove
the external batteries in order to prevent any damage that may be
caused by leakage.
The internal battery will need to be recharged and the device will
need to be reprogrammed before the device can be used on a patient.
➡
Insert freshly charged batteries.
➡
Set the switch in the internal battery compartment to ON and
leave the TM-2430 PC2 switched on for at least 2 hours. The inter-
nal battery will recharge during this time.
➡
Reprogramme the device.
➡
Replace the batteries with a set of freshly charged batteries
before fitting the device to a patient.
Charging batteries