64
Error messages
If any problems occur during the measurement, the display will show an error message
instead of a blood pressure reading.
Meaning of error messages:
Cause of
Err
message:
Troubleshooting:
Excessive variation in cuff pressure Hold arm still.
during measurement
No valid pulse readings
Check position of cuff (see page 53)
and repeat measurement.
Systole – diastole difference
Check position of cuff (see page 53)
too small:
and repeat measurement.
Systole – diastole
<
_
10 mmHg
Incorrectly inflated
Cuff possibly too loose, attach
cuff more firmly (see page 53)