96
13. Viewing the test results.
The log data that was obtained by the ATS2 macro must be evaluated by opening the Excel
sheet that was described in section “4-10. Pass/Fail evaluation and report of this test.”
When this sheet shows that the unit has passed test, it can be returned to the customer.
If the test is failed, retest or repair the ESP.
Print out the results of the Excel sheet and place it with the unit to be shipped back to the
customer. The passed Excel test sheet and text (.txt) log file should be archived so that there is
a record of the unit passing test.
TEST PROCEDURES
Device under test
According to OT1-400338
Serial Number
ESP-1
Lot
Date of test
2005,03,13
Tester's Name
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
AMPL
4.17
4.02
3.94
4.11
3.92
3.94
4.00
3.95
Nois
-80.03
-80.13
-80.05
-80.04
-81.30
-81.23
-81.30
-81.32
FRQ20K
3.39
3.08
3.05
3.15
2.87
2.98
2.74
3.04
FRQ20
3.66
3.50
3.41
3.61
3.44
3.47
3.54
3.48
THDN
0 0067
0 0068
0 0076
0 0075
0 0068
0 0068
0 0067
0 0067
Result
Passed
5-1.(GV=0dB)
Device under test
According to OT1-400338-1
Serial Number
Lot
Date of test
Tester's Name
CH1
CH2
CH3
CH4
CH5
CH6
CH7
CH8
AMPL
Nois
FRQ20K
FRQ20
THDN
Result
FAIL
5-1.(GV=0dB)
Содержание ControlSpace ESP-88
Страница 97: ...97 Figure 8 Motherboard Top Etch Layout Diagram Figure 9 Motherboard VCC Etch Layout Diagram...
Страница 98: ...98 Figure 10 Motherboard GND Etch Layout Diagram Figure 11 Motherboard Bottom Etch Layout Diagram...
Страница 100: ...100 Figure 14 MIC Line Input PCB GND Etch Layout Diagram Figure 15 MIC Line Input PCB Bottom Etch Layout Diagram...
Страница 101: ...101 Figure 16 DSP PCB Topside Etch Layout Diagram Figure 17 DSP PCB VCC Etch Layout Diagram...
Страница 102: ...102 Figure 18 DSP PCB GND Etch Layout Diagram Figure 19 DSP PCB Bottom Etch Layout Diagram...