ED01_REV01-2021 IFU OBWB_EN
5- 12
Danger:
Do not connect
the bracket to mains voltage.
Continuous current
1
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION
Read carefully
These operating instructions have been prepared using simple, easy-to-understand language. If
you have difficulty in interpreting what is written, contact the manufacturer for further
clarification.
Phone +39 0471 93 28 93
•
Please read these instructions carefully before using and installing the device.
•
The wall bracket is designed to allow the OB2012, OB1000 and OB3000 aspirators to be attached to emergency
vehicles, hospitals, clinics and/or other medical facilities.
•
The bracket cannot be used to fix other medical devices than those of the BSU family produced by Oscar Boscarol
srl.
•
The OBWB wall bracket, appropriately connected, with the supplied cable, to the external power source (SELV type
only) allows the internal battery to be recharged and the suction unit to be powered.
•
The bracket must be installed in accordance with the instructions in this manual and in compliance with the
requirements of EN 1789:2021. Failure to do so may jeopardise the safety of patients and users. The bracket is
tested to withstand positive and negative accelerations of up to 10 g.
•
Never alter the mechanical, electrical and structural parts of the bracket. Such interventions make the device
dangerous and do not allow its correct use, also damaging the fixed and powered device.
•
No technical intervention by the user is allowed on the device. The only operations allowed are those indicated in
these operating instructions. Any technical problems, periodic inspection and repairs should be referred to the
authorised service department and/or the manufacturer.
•
Always use only spare parts, supplied by the manufacturer (Oscar Boscarol srl), in order to ensure maximum
efficiency and safety of the device.
•
Do not remove and/or modify the coupling and release spring in order to ensure that the anchored device is held
up to negative acceleration forces of 10g.
If the user or patient becomes aware of a danger in use, a side effect, an accident caused
by the device or a critical issue (operational and design) not covered in these instructions
for use, he must immediately report this to the manufacturer at the following e-mail
address:
The bracket is designed in accordance with the requirements of EN 1789 for wall mounting
of the BSU Boscarol OB1000, OB2012 and OB3000 family of suction units. The power supply
is only for 12 VDC (OB2012-OB1000) and 11 to 30 VDC for the OB3000 suction unit. Never
connect the bracket to mains voltage or outside the range for SELV voltages.
2
IMPORTANT INFORMATION
The device has been designed and tested in accordance with the latest standards. Connecting the
device to vehicle electrical systems that are not compliant and/or not carried out by a professional
installer may damage the device and cause damage to the electrical system itself.