Pulling the suction unit towards you must not cause it to come loose. In the event of an accident, always have the wall
bracket and suction unit checked by the authorised service centre or manufacturer.
5.2
Cleaning the bracket
The bracket does not require any special cleaning, but disinfection may be necessary. In this case, proceed as follows:
•
Wear personal protective equipment (FFP2 masks, etc.).
•
Disconnect the fuse or the power cable of the bracket
•
Blow out the bracket with a compressed air source (not exceeding 2 bar) to remove any residue or scale.
•
Use non-aggressive disinfectants to clean the bracket (use disinfectants suitable for medical devices). Facilitate this
by using brushes or rags soaked in disinfectant.
•
When finished, dry all clean parts and reconnect the power supply.
•
Perform all tests described in chapter 4.6
In order to properly disinfect and decontaminate the device, we recommend that you purchase specific, approved
disinfectants for cleaning medical devices that are not harmful to humans or the environment. Do not use abrasive
substances.
Oscar Boscarol srl can supply you with such disinfectants (also suitable for our suction units). These disinfectants,
available in various formats (wipes, sprays, liquids) are laboratory tested and guarantee the deactivation of viruses,
bacteria and microorganisms. Used periodically, they destroy and prevent the formation of dangerous biofilms (surface
layers that easily harbour bacteria, moulds, viruses and microorganisms).
Deformation of the
electrical contacts
In the event of oxidation or alteration of the contacts, contact the authorised service
centre or the manufacturer.
For more detailed information, please contact us by email at
5.3
Checking the power supply
The wall bracket is also designed to allow powering and charging of the OB2012, OB1000 and OB3000 suction units. The
supply voltage must be between 12 and 15 Vdc for the OB2012 and OB1000 aspirators and between 11 and 30 Vdc for
the OB3000 aspirator. Any electrical checks on the vehicle's power supply system must always be carried out by
experienced and qualified personnel.
To check the voltage present on the contacts, use a voltmeter set for DC voltage range and measure the voltage present
on the two contacts of the bracket. The voltage measured must be as indicated in this user manual and in the following
figure. If there is no voltage or if there is a reversal of polarity, proceed as follows:
•
If the polarity is reversed, the device cannot be recharged and powered. Contact an authorised service centre
or the vehicle manufacturer.
•
If there is no voltage present, contact an authorised service centre or
reset the fuse.
•
If functional faults occur when switching on the device, contact a service
centre or the vehicle manufacturer.
5.4
Dismantling the wall bracket
When replacing or removing the wall bracket, always contact an authorised
service centre or the vehicle manufacturer.
5.5
Demolition and decommissioning
The device contains electrical parts that must be recycled according to the European Directive WEEE
2012/19/EU - Waste Electrical and Electronic Equipment and implemented in Italy by Legislative Decree
49/2014. The device also complies with Directive 2011/65/EC, which restricts and prohibits the use of
certain harmful substances in electronic and electrical equipment. No harmful substances that
violate the above-mentioned Directive are used in the production and assembly of the
electronic boards, in the wiring and in the connection of the electrical cables.
6
SPARE PARTS
There are no spare parts available for the device that can be bought and installed by users. You should always contact
the service centres or the manufacturer if you need to maintain or repair the device.
Positive pole (+)
Negative pole (-)