tipstim
®
Instruction Manual
19
Debugging and repair by untrained users
Untrained users must not attempt to repair or disassemble the device or its components/
accessories. Untrained users may carry out following tests:
Debugging and repair by authorized and qualifi ed personnel
Authorized personnel are allowed to carry out repairs of the device and its components/
accessories according to legal and technical standards. All persons wishing to be authorized
must to participate in a training course run by the manufacturer. The training course is
fee-paying.
No current
noticeable
The electrode cable is not
connected correctly to the
socket at the head end of
the device
Check the seating of the
plug at the device’s socket
error cause solution
Insuffi cient skin
contact
Replace
tipstim® glove
by
a new one. Replace surface
electrodes
Accessories are
damaged
Replace accessories
Current intensity settings
locked
Unlocking by pressing
“-“-button
Device cannot be switched
on
(Rechargeable) batteries
are completely dischar-
ged
Replace batteries
Device switches off imme-
diately after it has been
switched on
(Rechargeable) batteries
are completely dischar-
ged
Replace batteries
Reduced operating time
after recharge of batte-
ries
Rechargeable batteries are
worn out (more than 500
charging cycles)
Replace rechargeable
batteries
Current cannot be set
or display is blinking
(repeated acoustical
signal)