2
DEVICE DESCRIPTION
12
VTX3
www.bopp-reuther.de
12/2018 - Z-EN-09001-id-A Exi-Gb
Max. permissible product and ambient temperatures for devices with painted flow
sensors / signal converters or connection box mounted above the flow sensor
Temperature class in
°
C
Temperature class in
°
F
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
C
40
60
60
65
40
60
65
40
60
65
Nominal size
DN15
…
25
60
60
120
1
120
1
120
120
1
120
1
120
120
1
120
1
DN40
…
50
55
60
120
1
115
1
120
120
1
115
1
120
120
1
115
1
DN80
…
100
55
60
110
1
105
1
120
110
1
105
1
120
110
1
105
1
DN150...300
60
60
120
1
115
1
120
120
1
115
1
120
120
1
115
1
1
Permanent service temperature of connecting cable and cable entry min. 80
°
C
Temperature class
T6
T5
T4
T3
T2
T
amb
in
°
F
104
140
140
149
104
140
149
104
140
149
Nominal size
DN15
…
25
140
140
248
1
248
1
248
248
1
248
1
248
248
1
248
1
DN40
…
50
131
140
248
1
239
1
248
248
1
239
1
248
248
1
239
1
DN80
…
100
131
140
230
1
221
1
248
230
1
221
1
248
230
1
221
1
DN150...300
140
140
248
1
239
1
248
248
1
239
1
248
248
1
239
1
1
Permanent service temperature of connecting cable and cable entry min. 176
°
F