PMX40 RF Power Meter – INSTRUCTION MANUAL
3-3
current instrument state before shutdown. Pressing and holding
the On/Standby key for several seconds will force standby mode
if the instrument has become non-responsive. In this case, no
context save is performed.
10
Cooling Fan
Cooling air intake
11
GPIB
24-pin GPIB (IEEE-488) connector for connecting the power
meter to the remote control General Purpose Instrument Bus.
GPIB parameters can be configured through the menu.
12
AC Line Input
A multi-function power input module is used to house the AC
line input, main power switch, and safety fuse. The module
accepts a standard AC line cord, included with the power meter.
The power switch is used to shut off main instrument power.
The safety fuse may also be accessed once the line cord is
removed. The instrument’s power supply accepts 90 to
264 VAC, so no line voltage selection switch is necessary.
Caution
Replace fuse only with specified type and rating:
1.0A-T (time delay type), 250 VAC.
13
HDMI
HDMI receptacle for connecting an external monitor to mirror
front panel display. The image resolution will be 800 x 480 and
will be stretched to fit the external full display size.
14
LAN Ethernet
LAN connector for remote control and firmware updates. Allows
DHCP or fixed (IP / Subnet) setting mode. LAN parameters can
be configured through the menu.
15
Trig In
BNC input for connecting an external trigger signal to the power
meter. Voltage range is ±5 volts, but the input impedance is
1 Megohm to allow use of common 10x oscilloscope probe for a
±50 volt input range.
16
Multi I/O
BNC input/output for flexible use. May serve as a status or
alarm output, signal level monitor, or settable voltage source.