Before installation

 

 

 
 
 
 

 
Packaging Removal & Unit Placement

 

 
Before installation check the extraction unit for damage. 
All packaging must be removed before the unit is connected 
to the power supply.  

Please read all instructions in this manual before using 
this extractor. 

 

1. Move the unit to the location where it is going to be 
installed and remove the outer packaging. 

This unit 

should be installed in a well-ventilated area.

 

 
2. Open the front door. 
 Check the filters are located in their correct position 
before closing the door and securing the door latches. 

Note: The door will not close fully if the Combined filter 
has not been secured in place using the internal lever. 
(As detailed below) 

 
Ensure that 500 mm space is available around any vented 
panels on the extractor to ensure adequate airflow. 
 

Caution 

Due to the weight of the extractor suitable lifting 
equipment should be used and with regard to 

appropriate safety precautions. (See Appendix for 
product weight details) 

 

Caution 

Under no circumstances should the exhaust 

outlet/s be covered as this will restrict the airflow 
and cause overheating. 
 

Caution 

Do not block or cover the cooling vents on the unit, 

as this severely restricts airflow and may cause 
damage to the unit. 

 
 
 
 
 

 

 

 

 

 

 

 

 

 

The FumeCab 700 features a hinged transparent Perspex 
panel on the front of the unit. This allows for more room to 
place larger work pieces inside the unit when working, while 
not restricting the user’s vision. 

 

 

 

 

 

01 

Содержание FumeCAB 700

Страница 1: ...FumeCAB 700 115V ...

Страница 2: ...1 Overview of your extraction system front Overview of your extraction system back Overview of Control Panel Important safety notes Safety labels Unpacking and unit placement Unit Information Connecting to Mains Turning extractor On changing display units Cleaning the unit Pre filter replacement Combined filter replacement Consumable spares Filter disposal FumeCab 700 Specifications ...

Страница 3: ...Overview 1 01 Company Branding Control Panel Filter Door Panel Hinged Perspex enclosure panel Door Latch Combined Filter Filter Release Handle Pre Filter Air Inlet Mains Isolation Switch ...

Страница 4: ...Overview 1 02 Motor Cooling Mains Inlet Inlet for Tools Equipment Cables Air outlet ...

Страница 5: ...Overview 1 03 Main Isolation Switch Lighting Switch Filter Change Indicator Company Branding ...

Страница 6: ...oving the pump motor access panel 2 Use only as described in this manual 3 Connect the system to a properly grounded outlet Dangers to eyes breathing and skin Once used the filters within the FumeCab 700 system may contain a mixture of particulates some of which may be sub micron size When the used filters are moved it may agitate some of this particulate which could get into the breathing zone an...

Страница 7: ...f Manufacture Relevant approval markings logos PLEASE NOTE If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment maybe compromised Fire Risk Warning In the very rare event that a burning ember or spark is drawn into the fume extraction unit it may be possible that the filters will ignite Whilst any resultant fire would typically be retained...

Страница 8: ...been secured in place using the internal lever As detailed below Ensure that 500 mm space is available around any vented panels on the extractor to ensure adequate airflow Caution Due to the weight of the extractor suitable lifting equipment should be used and with regard to appropriate safety precautions See Appendix for product weight details Caution Under no circumstances should the exhaust out...

Страница 9: ...ns The supply cord should only be replaced by a BOFA engineer as an electrical safety test may be required after replacement The extraction unit MUST be connected to a properly earthed outlet Connect the power cable to an isolated electrical supply The mains socket should be installed near the extractor it should be easily accessible and able to be switched On Off The cable run should be arranged ...

Страница 10: ... switched to the On position the top side of the switch should be depressed to power the unit Please Note To ensure efficient airflow the FumeCab must be used with the clear panel in the down position The extractor and all pipe work must be fully installed and connected before the airflow is set 5 01 ...

Страница 11: ...vent build up of dust and overheating of the unit Replacing Filters The FumeCab 700 system will alert the user when its filters need replacing when the red filter change indicator light on the control panel glows Fire Risk Warning In the very rare event that a burning ember or spark is drawn into the fume extraction unit it may be possible that the filters will ignite Whilst any resultant fire wou...

Страница 12: ...d filter change indicator located on the control panel glows To remove and replace the combined filter follow the procedure detailed below 1 Isolate the electrical supply to the extractor 2 Undo the catch on the front of the unit and open the filter door 3 Rotate the lever below the filter through 180o to lower the combined filter 4 Using the handle on the front of the filter pull it out of the un...

Страница 13: ...filter Date Engineer Date Engineer Filter disposal The Pre and Combined filters are manufactured from non toxic materials Filters are not re usable cleaning used filters is not recommended The method of disposal of the used filters depends on the material deposited on them For your guidance Deposit EWC Listing Comment Non Hazardous 15 02 03 Can be disposed of as non hazardous waste Hazardous 15 02...

Страница 14: ...ing speed Size Metric mm Imperial inches Height 780 30 7 Depth 490 19 3 Width 690 27 1 Filters Filter Type Surface area Efficiency Pre filter 1 68 m2 99 0 3micron Combined filter 3 0m2 99 997 0 3micron Combined Filter Gas section Filter Type Carbon type Amount Combined filter Gas Activated carbon 17kgs Environmental operating range Temperature 5o C to 40o C Humidity Max 80 RH up to 31o C Max 50 RH...

Страница 15: ... Information BOFA Headquarters 21 22 Balena Close Creekmoor industrial Estate Poole Dorset BH17 7DX UK Phone 44 0 1202 699444 BOFA Americas 303 S Madison Street Staunton Illnois 62088 USA Phone 001 618 205 5007 ...

Отзывы: