Maintenance 

 

 
 
 
 
 
 
Maintenance UK 

It is a legal requirement, under regulation 9 of the COSHH 
regulations that all local exhaust ventilation systems are 
thoroughly examined and tested at least once every 14 
months (typically carried out annually). The approved code 
of practice recommends that a visual check should be 
carried out at least once a week. 
COSHH requires the annual inspection and testing to be 
carried out by a competent person and specifies that 
documentation results are recorded in a log.  
Contact the seller for more information about inspection 
and certification.  
 

Maintenance General 

User maintenance is limited to cleaning the unit and filter 
replacement, only the manufacturers trained maintenance 
technicians are authorised to carry out component testing 
and replacement. Unauthorised work or the use of 
unauthorised replacement filters may result in a potentially 
dangerous situation and/or damage to the extractor unit 
and will invalidate the manufacturer’s warranty. 
 

Cleaning the unit

 

The powder coat finish units can be cleaned with a damp 
cloth and non-aggressive detergent, do not use an abrasive 
cleaning product as this will damage the finish. 
 
The cooling inlets and outlets should be cleaned once a year 
to prevent build-up of dust and overheating of the unit.  
 

Replacing Filters 

The FumeCab 700 system will alert the user when its filters 
need replacing when the red filter change indicator light on 
the control panel glows. 

Fire Risk Warning                                                                                  

In the very rare event that a burning ember or spark is drawn 
into the fume extraction unit, it may be possible that the 
filters will ignite. Whilst any resultant fire would typically be 
retained within the fume extraction unit, the damage to the 
extractor would be significant. It is therefore essential to 

minimise this risk by undertaking an appropriate 
assessment to determine:-

 

a). Whether additional fire protection equipment should be 
installed.  

b). Appropriate maintenance procedures to prevent the risk 
of build-up of debris which could potentially combust. 
                    

                          
A log of the changes should be maintained by the user. 
The filters require attention when the display shows the 
filter blocked icon/ filter output signal (if fitted) or when the 
extractor no longer removes fume efficiently. 
 
All filters are tested to BS3928. A certificate of conformity 
for each filter is available on request. 
It is recommended that a spare set of filters are kept on site 
to avoid prolonged unit unavailability. Part numbers for 
replacement filters can be found on the filters fitted in your 
system.  
To prevent overheating, units should not be run with a 
blocked filter condition, or with dust obstruction of Inlets / 
Outlets. 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 

 

01 

 

Содержание FumeCAB 700

Страница 1: ...FumeCAB 700 115V ...

Страница 2: ...1 Overview of your extraction system front Overview of your extraction system back Overview of Control Panel Important safety notes Safety labels Unpacking and unit placement Unit Information Connecting to Mains Turning extractor On changing display units Cleaning the unit Pre filter replacement Combined filter replacement Consumable spares Filter disposal FumeCab 700 Specifications ...

Страница 3: ...Overview 1 01 Company Branding Control Panel Filter Door Panel Hinged Perspex enclosure panel Door Latch Combined Filter Filter Release Handle Pre Filter Air Inlet Mains Isolation Switch ...

Страница 4: ...Overview 1 02 Motor Cooling Mains Inlet Inlet for Tools Equipment Cables Air outlet ...

Страница 5: ...Overview 1 03 Main Isolation Switch Lighting Switch Filter Change Indicator Company Branding ...

Страница 6: ...oving the pump motor access panel 2 Use only as described in this manual 3 Connect the system to a properly grounded outlet Dangers to eyes breathing and skin Once used the filters within the FumeCab 700 system may contain a mixture of particulates some of which may be sub micron size When the used filters are moved it may agitate some of this particulate which could get into the breathing zone an...

Страница 7: ...f Manufacture Relevant approval markings logos PLEASE NOTE If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment maybe compromised Fire Risk Warning In the very rare event that a burning ember or spark is drawn into the fume extraction unit it may be possible that the filters will ignite Whilst any resultant fire would typically be retained...

Страница 8: ...been secured in place using the internal lever As detailed below Ensure that 500 mm space is available around any vented panels on the extractor to ensure adequate airflow Caution Due to the weight of the extractor suitable lifting equipment should be used and with regard to appropriate safety precautions See Appendix for product weight details Caution Under no circumstances should the exhaust out...

Страница 9: ...ns The supply cord should only be replaced by a BOFA engineer as an electrical safety test may be required after replacement The extraction unit MUST be connected to a properly earthed outlet Connect the power cable to an isolated electrical supply The mains socket should be installed near the extractor it should be easily accessible and able to be switched On Off The cable run should be arranged ...

Страница 10: ... switched to the On position the top side of the switch should be depressed to power the unit Please Note To ensure efficient airflow the FumeCab must be used with the clear panel in the down position The extractor and all pipe work must be fully installed and connected before the airflow is set 5 01 ...

Страница 11: ...vent build up of dust and overheating of the unit Replacing Filters The FumeCab 700 system will alert the user when its filters need replacing when the red filter change indicator light on the control panel glows Fire Risk Warning In the very rare event that a burning ember or spark is drawn into the fume extraction unit it may be possible that the filters will ignite Whilst any resultant fire wou...

Страница 12: ...d filter change indicator located on the control panel glows To remove and replace the combined filter follow the procedure detailed below 1 Isolate the electrical supply to the extractor 2 Undo the catch on the front of the unit and open the filter door 3 Rotate the lever below the filter through 180o to lower the combined filter 4 Using the handle on the front of the filter pull it out of the un...

Страница 13: ...filter Date Engineer Date Engineer Filter disposal The Pre and Combined filters are manufactured from non toxic materials Filters are not re usable cleaning used filters is not recommended The method of disposal of the used filters depends on the material deposited on them For your guidance Deposit EWC Listing Comment Non Hazardous 15 02 03 Can be disposed of as non hazardous waste Hazardous 15 02...

Страница 14: ...ing speed Size Metric mm Imperial inches Height 780 30 7 Depth 490 19 3 Width 690 27 1 Filters Filter Type Surface area Efficiency Pre filter 1 68 m2 99 0 3micron Combined filter 3 0m2 99 997 0 3micron Combined Filter Gas section Filter Type Carbon type Amount Combined filter Gas Activated carbon 17kgs Environmental operating range Temperature 5o C to 40o C Humidity Max 80 RH up to 31o C Max 50 RH...

Страница 15: ... Information BOFA Headquarters 21 22 Balena Close Creekmoor industrial Estate Poole Dorset BH17 7DX UK Phone 44 0 1202 699444 BOFA Americas 303 S Madison Street Staunton Illnois 62088 USA Phone 001 618 205 5007 ...

Отзывы: