2-2-2 Installing the Trinocular
Viewing Unit
Insert the trinocular viewing unit (figure4)
into the microscope head (figure5), turn to
a proper position, then use the hexagon
wrench screw down the bolt to fix ( See
figure 5 ).
2-2-3 Installing and Replacing the
Lamp (figure 6)
Please use the specified halogen Lamp
6V30W.
1
.
Hold to the bulb after you wrap it with
gauze or other protection materials, and
then deeply insert it into the lamp holder.
2. Replacing Lamp when using or soon after
When using, or soon after it is turned off,
the lamp, the lamp house and nearby
parts will be very hot and will cause
serious burns. Please turn the main
switch on “O” (off), pull out power plug,
and make sure the bulb, the lamp room
and periphery are all cool. Then, you can
do your replacing.
Please insert the lamp gently, or it will be
damaged by excessive extrusion
Do not touch the Halogen bulb with
your bare hands. It will shorten the
service life or cause it to burst. If you
leave finger marks on the surface
carelessly, clean it with a dry soft cloth.
4
5
Bolt
6
Bolt
Lamp House Bolt