STP-603/1003 Series Instruction Manual
6.2.3 Installation
Area
Installation
Leave enough space for the following in addition to that for the STP control unit
•
Space for maintenance and inspection
•
Space for inlet and outlet of air for cooling
- Top and side: 5 cm or more
- Bottom:
1.8 cm or more (height of the rubber feet)
•
Space for connecting the cables
- Rear:
11.5 cm or more
◇
The minimum bending radius of the STP connection cable is
100 mm (See Figure13.4, "External Appearance of the STP Control
Unit" [Cable Space]).
DO NOT excessively bend the cables and beware of any obstacles
when installing the STP pump.
Also, leave enough space to install other cables without bending
them excessively.
5cm
1.8
c
m
or
m
o
re
11.5cm
or more
5cm
or more
or more
5cm
or more
or
m
or
e
5c
m
Figure 6.3 Peripheral Space of the STP Control Unit
6-6