Issue A
Nov 02
PAGE
18
XDS Dry Pump
OPERATION
18
4.7
Shut down
Use the procedure below to shut down the pump:
1.
If shutting the pump down prior to a period of
storage, remove any process gases by running
on high flow gas-ballast for at least one hour.
2.
Close any vacuum system isolation valve to
prevent suckback into the vacuum system
(where fitted).
3.
Switch off the pump using the on/off switch.
When the pump is switched off the fan will
continue to run for 1 minute.
4.
Vent the XDS35i by the gas ballast control or
valve on the inlet.