8-41
575 DISPlay Commands (SCPI Command Summary)
Keyword
Parameter
Comments
:DISPlay
Subsystem. Contains commands to control
the display.
:MODe
0/1 or ON/OFF
Enables/Disables automatic display update.
When true, front panel display is updated
with serial command parameter changes.
Setting to false decreases response time.
:UPDate?
Query only. Forces update of display. Use
when mode is false.
:BRIGhtness
0-4
Controls intensity of display. Range is 0 to 4,
where 0 is off and 4 is full intensity.
:ENABle
0/1 or ON/OFF
Enables/Disables the display and front panel
lights. When Disabled the keylock is enabled
to prevent parameter changes from the front
panel.
IEEE 488.2 Common Commands
Keyword
Parameter
Comments
*IDN?
Identification
Query
Queries the Pulse Generator Identification.
The ID will be in the following format:
manufacturer,model#,serial#,version#
*RCL
0-12
Restores the state of the Pulse Generator
from a copy stored in local nonvolatile
memory (0 through 12 are valid memory
blocks).
*RST
Reset Command
Resets the Pulse Generator to the default
state.
*SAV
1-12
Stores the current state of the Pulse
Generator in local nonvolatile memory (1
through 12 are valid memory blocks).
*TRG
Trigger
Generates a software trigger pulse.
Operation is the same as receiving an
external trigger pulse.
*LBL
Setup Label
Query Form returns the label of the last
saved or recalled configuration.
Command Form sets the label string for the
next "*SAV" command. String must be in
double quotes, 14 characters max.
*ARM
Channel Trigger
Reset
Resets channel triggers when channels are
set to single shot or burst mode. Functions
like pressing the function then run/stop
button.