Cleaning and Lubrication 

 

Protect the tester with dust-proof cover when it will not be used for a longer period of time.   

 

Fill adequate machine oil to the contacts between lead screw and hand wheel periodically.   

 

The friction surfaces should be lubricated regularly, 2 times per month.   

 

The decelerator should be filled with grease #20 once every half year. (industrial Vaseline) 

14

 

Cautions for operation 

Fitting Indenter: All the indenters are composed of fixing rod, fixing nut and tungsten carbide ball (Fig-2). Before 

fitting indenters, the ball groove of fixing rod and the carbide must be cleaned with acid-free gasoline and make it 

dry. Paint some acid-free Vaseline into the groove, fit the carbide ball and tighten the fixing nut. During the testing, 

check occasionally whether the carbide ball is loose in the indenter. In case the ball is loose, the test is invalid. 

During the process of loading, dwelling and unloading, never turn the rotation turret. Otherwise, the indenter may 

be damaged. 

Selection  of  power  voltage:  the  voltage  should  be  220V  (or  110V,  380V  according  to  request)  and  must  be 

grounded properly.   

Temperature of specimen: The temperatures of specimen shall be kept at  normal atmospheric temperature  during 

the testing. In other special circumstances, the temperature of ferrous metals should not exceed 100

, but that of 

non-ferrous metals must be kept under ambient temperature. 

The  diameter  of  tungsten  carbide  ball  shall  be  inspected  regularly  in  case  of  exceeding  its  tolerance.  When  the 

diameter is 2.5mm, the tolerance should not be more than 

±

0.003mm. When the diameter is 5mm, the tolerance 

should be less than 

±

0.004mm. When the diameter is 10mm, the tolerance should be less than 

±

0.005mm the 

surface finish Rz should be no more than 0.2 micron. 

Distance between the surface of specimen and the indenter: It is recommended that 0.2mm to 0.4mm be controlled 

at  the  beginning.  After  certain  period  of  repeated  operation,  when  the  operator  is  able  to  judge  the  position  of 

specimen through the image of ocular lens, the 2X objective lens can be used.   

15

 

Troubleshooting 

    In case  of breakdown, the following may be of help to find out the reason of breakdowns and recommend the 

remedies. If not successful, contact our aftersales service department rather than dismantle the tester by yourself.   

 

Table 7 Troubleshooting Guideline 

Breakdowns 

Reasons 

Remedies 

Screw rod not 

elevating fluently 

Rusty or chips 

Remove the chips around lead screw and hand wheel, and then fill some 

lubricants 

Test results not 

precise 

Distance of indentations 

The centers of two adjacent indentations are too close to each other 

Indenter 

Replace the indenter if damaged 

Worktable 

Check whether there are foreign materials at the contacts of work table and 

lead screw 

Check whether the selected anvil is suitable for the specimen or not 

The protective sleeve of lead screw has propped up the work table 

Specimen 

 

Test force is applied to the specimen perpendicularly or not 

Whether there are burrs on the back 

Too thin 

Surroundings 

Mechanical vibration around 

Ambient temperature kept at 10-30

º

Others 

Calibrate with standard hardness block 

 

Содержание 3000-OBC

Страница 1: ... Makina Sanayi Ve Ticaret Ltd Şti İkitelli Organize Sanayi Bölgesi Dolapdere Sanayi Sitesi Ada 4 No 7 9 Başakşehir İSTANBUL TURKEY Phone 90 212 671 02 24 671 02 25 Fax 90 212 671 02 26 web www bulutmak com e mail bms bulutmak com OPERATIONAL MANUAL ...

Страница 2: ...ptions of Mechanism Parts 5 6 Preparation 6 7 Package Opening 6 8 Cautions of Transportation 7 9 Installation and Commissioning 7 10 Preparation prior to normal operation 8 11 Operation of micrometer 12 12 Testing procedures 13 13 Cleaning and Lubrication 14 14 Cautions for operation 14 15 Troubleshooting 14 ...

Страница 3: ...ns Symbols Descriptions Unit D Diameter of tungsten carbide ball mm F Test forces N d Average diameter of indentation mm d1 d2 Diameters of indentations along two perpendicular directions mm h Depth of indentation mm HBW Brinell hardness HBW Fig 1 Testing Principle diagram of Brinell hardness To apply specified test forces onto the surface of test specimen through a tungsten carbide ball of certai...

Страница 4: ... gray cast iron 10 1000 250 62 5 15 6 10 0 48 315 5 500 125 31 2 7 8 5 0 23 8 158 Annealed aluminum 2 5 250 62 5 15 6 3 9 2 5 11 9 79 Bearing alloy 1 25 125 31 2 7 8 2 0 1 2 6 0 39 Conductor wires 0 5 50 12 5 3 1 0 8 0 5 2 4 15 8 Soft materials 4 Technical Parameters Main Technical Parameters are included in Fig 3 and Table 3 Table 3 Main Technical Parameters Item code Content Measuring range 8 65...

Страница 5: ...utomatically by the loading and unloading parts which are fitted inside the machine body The rotation of motor is transferred to the screw rod When the screw rod descends the test force is passed onto the indenter via larger lever 12 smaller lever 17 and main shaft Main shaft parts are composed of main shaft bench main shaft indenter seating rotary parts seating rotary part connection nut and conn...

Страница 6: ...ove the wood screw fastening the packing box and the bottom with pliers and lift up the packing box 3 Take out the document bag and verify that it is complete with Operation Instruction Packing List and Certificate of Quality 4 Loosen the nuts of hooks fastening the accessory box with a spanner and remove the pressure plate accessory box Check the parts in accordance with the Packing List 5 Remove...

Страница 7: ...efore installation get ready the spanner level meter and larger anvil 1 Place the tester on the prefabricated work table in the way that the screw rod falls into the hole 2 Remove cushion block of indenter Fig 6 3 Removing wood cushion of the indenter Turn the hand wheel anticlockwise so that the screw rod descends and the wood cushion can be removed therein Should the screw rod fail to descend wh...

Страница 8: ...oot power socket is necessary and it should be well grounded Take out the printer and data lines and plug properly Take out the weights and hang them onto the larger lever Fig 7 Removing wood wedge of larger lever Fig 6 Removing wood cushion of the indente 10 Preparation prior to normal operation 1 Prefabrication of specimen the specimen shall be polished with smooth and flat surface free from lay...

Страница 9: ... 4 seconds to 99 seconds D The specimen shall meet the following requirements 1 The specimen shall be polished so that the edges of indentation are visible enough to ensure the precision of indentation measurement Also its surface shall be free from oxide skins electroplated layers oil dirt and any surface defects 2 The surface to be tested shall be parallel to the supporting face so that the test...

Страница 10: ... the ocular lens that the two division lines are in parallel without any space between them In case that the two division lines are not clearly visible turn the ocular lens until they are clear and then press ZERO to make it zero The above operation must be repeated whenever the power is turned off and on 3 Display and Control panel Press MENU key then the interface will show as Fig 9 Fig 9 Menu i...

Страница 11: ... page press ENTER to enter the menu shown in Fig 10 11 When the cursor comes to Quit to press ENTER to exit the preset menu and return to the interface of Fig 8 Fig 10 Menu interface 12 When the cursor comes to the item Test points to press ENTER to preset whether or not to carry out data statistics whose range is from 1 to 99 If 1 is selected no statistics will be done i e to print out the test r...

Страница 12: ... to is including C_St Cr_St CrV_St CrNi St CrMo_St CrNiMo_St CrMnSi_St Strength_St and Stainless_St Press or key to select specific item then press ENTER and press or key to change the figure At last press ENTER again to confirm 11 Operation of micrometer Turn the 1X or 2X objective lens to the front and then turn elevation hand wheel so that the surface of specimen is 15mm under the objective len...

Страница 13: ...orting surface of specimen and put it onto the worktable Turn the hand wheel to move the worktable up slowly until the specimen is 0 2 mm to 0 4mm away from the indenter or the surface of specimen can be observed clearly from the 2X objective lens It is recommended that 0 2mm to 0 4mm be controlled at the beginning After certain period of repeated operation when the operator is able to judge the p...

Страница 14: ...ase of exceeding its tolerance When the diameter is 2 5mm the tolerance should not be more than 0 003mm When the diameter is 5mm the tolerance should be less than 0 004mm When the diameter is 10mm the tolerance should be less than 0 005mm the surface finish Rz should be no more than 0 2 micron Distance between the surface of specimen and the indenter It is recommended that 0 2mm to 0 4mm be contro...

Отзывы: