  Transport Lock 

4.1

When the SPECTROstar Nano microplate reader is shipped, the transport lock for the plate carrier is in the 
locked position and must be unlocked prior to using the instrument. Once the instrument is placed at its 
designated lab bench position unlock the plate carrier.   

 

 

If the SPECTROstar Nano  has to be moved or shipped to another location, it is  very 
important to lock and secure the transport lock in order to avoid damage to the microplate 
carrier system 

 

 

 
Unlocking the Plate Carrier 

 
The transport pin that locks the plate carrier can be accessed from the top of the instrument. Locate 
the silver screw located next to the label “Transport Lock” (figure 5). Use the supplied 5,5 mm slotted 
screwdriver and unlock the pin by turning the screw counter clockwise. The plate carrier is unlocked if 
the screw moves freely up and down. Once the SPECTROstar Nano has been turned on, press the 
In/Out button (figure 10) briefly and the plate carrier should move out.  
  
 

    
 
 
 
 
 
 
 
 
 
 
 
 
 
  
 

 
Figure 5: Transport Lock position 
 
 
 

Locking the Plate Carrier 

 
Upon closing the Control Software or pushing the plate carrier In/Out button for more than three seconds 
(figure 10), the plate carrier will move into the respective lock position. Slowly and carefully push the screw 
down with the slotted 5,5 mm screwdriver and turn it clockwise to fasten and to lock the plate carrier. The 
screw has to be tight and must not be moving as it was in the unlocked position.  

 

 

 

 

 
 
   
 
  
 
 
 
 
 
 
 
 

 

Содержание spectrostar nano

Страница 1: ...SPECTROstar Nano Operating Manual Revision C...

Страница 2: ...formats By using proprietary technology in a variety of important parts the SPECTROstar Nano delivers high end results SPECTROstar Nano Manufacturer BMG LABTECH GmbH Allmendgr n 8 77799 Ortenberg Ger...

Страница 3: ...ation is subject to change without notice BMG LABTECH GmbH shall not be liable to any extent whatsoever for any damages resulting from or arising out of the use or performance of this system and relat...

Страница 4: ...6 CLEANING AND INSTRUMENT DISINFECTION 9 2 7 DISPOSAL AND INSTRUMENT TAKEBACK RECYCLING PROGRAM 11 2 7 1 DISPOSAL OF TRANSPORT PACKING 11 2 7 2 WEEE COMPLIANCE 11 3 OVERVIEW 12 3 1 FRONT AND KEYPAD C...

Страница 5: ...m height 8 5 mm Light Source High energy xenon flash lamp Detector Spectrometer with CCD Spectral Range 220 1000 nm for ABS Absorbance Spectrometer Sensitivity ABS with Spectrometer Selectable Spectra...

Страница 6: ...conformity Over voltage category II Contamination class II Protection class III Dimensions Width 36 cm depth 50 cm height 16 cm Weight 11 kg Options Gas Vent system to inject an atmosphere or to pull...

Страница 7: ...an lead to unsafe operation of this equipment The area designated for the instrument should be free of dust liquids and acidic vapour The surface of the table should be flat and even Avoid areas subje...

Страница 8: ...jure the eye Thus avoid looking at the light directly During normal operation the user will not encounter any light from the xenon flash lamp Chemical and Biological Safety 2 5 Daily routine with this...

Страница 9: ...authorities Use suitable disinfectants e g Alcohol 70 Authorized personnel wearing disposable gloves and protective clothing should only perform the disinfection procedure The location should be well...

Страница 10: ...the instrument and its inventory have been disinfected according to the instructions given in the Operating Manual Name ________________________________________________ Firm __________________________...

Страница 11: ...involved in the life cycle of electrical and electronic equipment The directive specifies that as of August 13 2005 manufacturers of specified electronic equipment will be obliged to take back these p...

Страница 12: ...TECH s LVis Plate for low volume measurements down to 2 L and for instrument performance testing 3 1 2 Cuvette Port The SPECTROstar Nano comes with an integrated cuvette port for standard cuvettes wit...

Страница 13: ...red reader error has occurred MICROPLATE In Out Reader is in standby Short click microplate carrier will move In Out Long click 3 sec microplate carrier will move to transport lock position MICROPLATE...

Страница 14: ...ument Control and MARS Data Analysis software CD ROM in a cover inside the manual Manual also on CD ROM Power cord External Power Supply USB cable Slotted 5 5 mm screwdriver Call BMG LABTECH immediate...

Страница 15: ...p of the instrument Locate the silver screw located next to the label Transport Lock figure 5 Use the supplied 5 5 mm slotted screwdriver and unlock the pin by turning the screw counter clockwise The...

Страница 16: ...n and power up the PC laptop The Power LED shows a steady green light indicating that the SPECTROstar Nano is now turned On and in standby Double click the SPECTROstar Nano icon A window will appear a...

Отзывы: