BMC Polymate YH-1000C Скачать руководство пользователя страница 29

Polymate

®

 YH-1000C Polysomnograph User Manual V3.1

 

25 

 

 

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity 

The Polysomnograph (YH-1000C) is intended for use in the electromagnetic environment 
specified below. The customer or the user of the Polysomnograph (YH-1000C) should assure that 
it is used in such an environment. 

Immunity Test 

IEC 60601 

Test Level 

Compliance 
Level 

Electromagnetic 

Environment - Guidance

Electrostatic Discharge (ESD) 

IEC 61000-4-2 

±6 kV 
Contact 

±8 kV Air 

±6 kV Contact 

±8 kV Air 

Floors should be wood, 
concrete or ceramic tile. If 
floors are covered with 
synthetic material, the 
relative humidity should be 
at least 30%. 

Power Frequency (50/60 Hz) 
Magnetic Field 

IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical home 
or hospital environment. 

Содержание Polymate YH-1000C

Страница 1: ...Polymate YH 1000C Polysomnograph User Manual 0123...

Страница 2: ...sors to the Recorder 10 4 1 3 Applying the Sensors to Patient 11 4 2 Instructions for Recorder 13 4 2 1 Turn on off the Recorder 14 4 2 2 Recorder Information 14 4 2 3 Stop Record 18 CHAPTER 5 CLEANIN...

Страница 3: ...nstall Analysis Software 28 10 1 2 Install the USB driver 32 10 2 Collaborated Work and Analysis Data 38 10 2 1 Data reception 38 10 2 2 Data file management 41 10 2 3 Data playback 42 10 2 4 Print pr...

Страница 4: ...of YH 1000C Polysomnograph This manual is only used by professional sleep lab or physician Please read this manual before using the device to make sure operating safely Copyright The copyright of thi...

Страница 5: ...icles If any of these items are not included contact a BMC representative No Articles Num N O 1 Recorder Host 1 N 2 Oral and Nasal Airflow Cannula 1 N 3 SpO2 Sensor 1 N 4 Body Position Snoring Chest a...

Страница 6: ...ate to the hose PC CPAP pressure cannula allow the Ploysomnograph to record data from these pressure therapy device RESmart CPAP or APAP Oxygen saturation and pulse measurements are made by a SpO2 sen...

Страница 7: ...fabric package on the patient s abdomen or chest area The data collected by the recorder is analyzed using the PolyLogic Analysis software for Polysomnograph 1 3 Symbols Type BF Applied part Operatin...

Страница 8: ...snoring event auto analysis z Editing z Reporting and printing 2 2 Warnings A Warning indicates the possibility of injury to the patient or operator z Do not use the Polymate recorder for continuous...

Страница 9: ...passed through the tissue Low perfusion caused by blood pressure cuff arterial or intravascular catheter or poor blood flow Portable and mobile Radio Frequency communications equipment Electromagneti...

Страница 10: ...otential safety hazard z The oral and nasal airflow cannula is a disposable component and designed for single patient use only z The pulse oximeter probe which is used for adult 40kg only may be used...

Страница 11: ...tructions and in accordance with local regulations BMC is not responsible for the disposal of the 1 5V alkaline battery 2 4 Contraindications z Users allergic to the device z Users planted with cardia...

Страница 12: ...choose a variety of functions on the LCD screen to achieve the various operations See the following Instructions for details There are three sensor inputs at the recorder bottom The flow input is use...

Страница 13: ...ond can turn on off the recorder which will not respond to a short time click The small button is used to reset the recorder in unusual situation Usually it is unnecessary to be used so it only can be...

Страница 14: ...efore inserting the battery The recorder may be damaged if the battery is inserted incorrectly When the battery voltage is lower than 2 3 volts the recorder can t work 4 1 2 Connecting of the Sensors...

Страница 15: ...the same color regardless of positive or negative electrode can be swapped each other To obtain the accurate data BMC suggest using single use ECG electrode White One palm distance left from navel Gr...

Страница 16: ...Then fix sensor with plaster as shown in Figure 4 In order not to affect the patient s action it is suggested that the sensor fixed on the patient s forefinger at the same time the sensor cable should...

Страница 17: ...6 For the purpose of pressure therapy with the Polymate please select the therapy device specified by BMC Figure 6 Fixed Position of CPAP Pressure Cannula 4 2 Instructions for Recorder Explanation th...

Страница 18: ...e operation is same as turn on the recorder 4 2 2 Recorder Information After turn on the main interface appears as follows Figure 7 Recorder main interface Recorder operations and functions choose can...

Страница 19: ...battery voltage Because of the different type of batteries when the left battery power is the same the battery voltage may be different If replacement of other type batteries within this function by s...

Страница 20: ...y prior to the completion of the erase Note Electric current when erase is much larger than normal status Low battery can lead to the failure of data erase please notice the tips on the LCD for the er...

Страница 21: ...n of record can be observed in ever changing Before the start of record pressing Shift Cancel button can withdraw from the record screen and return to the main interface Once the record has already be...

Страница 22: ...ue the countdown If the boot time exceed to the preset time of the timing record the recorder will immediately begin to record as soon as the recorder is turned on Note DO NOT remove the battery after...

Страница 23: ...clip probe and LEDs to remove oil and dirt 3 Wipe the outside of the clip with a damp cloth 4 Dry the sensor with a dry cloth Cleaning the Chest and Abdomen Effort Sensor 1 Disconnect the sensors fro...

Страница 24: ...ct Sterilization z If necessary i e after infectious diseases or if parts have become unusually dirty you can disinfect certain parts by wiping them Follow the instructions for the disinfectant used W...

Страница 25: ...ng no signals Plug connections are not properly connected Ensure the connections are firm Display too dim Display defective Return the device to BMC for repair Display dark Batteries rechargeable batt...

Страница 26: ...Hz So even there are some filtering methods to process the SpO2 data it will not delay for a relative long time to the displaying data Signal Processing z Pulse Oximetry Calculation z Pulse Rate Calcu...

Страница 27: ...0Hz 300Hz Power Internally Powered Equipment Battery Type 2X1 5V R6 AA Battery Duration 17 hours minimum Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Elect...

Страница 28: ...romagnetic Environment Guidance RF Emissions CISPR 11 Group 1 This Device uses RF energy only for its internal function Therefore its RF radiated emissions are very low and are not likely to cause any...

Страница 29: ...ironment Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance Electrostatic Discharge ESD IEC 61000 4 2 6 kV Contact 8 kV Air 6 kV Contact 8 kV Air Floors should be...

Страница 30: ...ce in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in...

Страница 31: ...sed by accident misuse abuse alteration and other defects not related to material or workmanship Damages owing to improper operations will ask for paying for the cost fee when servicing BMC MEDICAL CO...

Страница 32: ...10 1 1 Install Analysis Software z Insert Analysis Software disk to your disk drive z The Installation programs will automatically startup within 10 seconds The executable program SETUP exe inside the...

Страница 33: ...Polymate YH 1000C Polysomnograph User Manual V3 1 29 a Click Next to enter the user information input page b Input user information and click Next...

Страница 34: ...Polymate YH 1000C Polysomnograph User Manual V3 1 30 c Select the installation location and click Next The default directory is C BMC Polysomnograph d Check current settings and click Next...

Страница 35: ...Polymate YH 1000C Polysomnograph User Manual V3 1 31 e Check current settings and click Next f Wait for software installation and icon creation to complete automatically...

Страница 36: ...icon on desktop to start the software 10 1 2 Install the USB driver z This driver is intended for PC to identify the recorder You can continue to install it following the above processes Or double cli...

Страница 37: ...Polymate YH 1000C Polysomnograph User Manual V3 1 33 b Wait while system is scanned and driver files are copied c During this process a prompt box may pop up and click Continue Anyway...

Страница 38: ...ing the USB cable When the system finds the hardware the Found New Hardware Wizard pops up to set up the new device Select the option No not this time when asked Can Windows connect to Windows Update...

Страница 39: ...mate YH 1000C Polysomnograph User Manual V3 1 35 c Set the options according to the figure in the search and installation options page Click Browse button to add the directory of the driver and click...

Страница 40: ...Polymate YH 1000C Polysomnograph User Manual V3 1 36 d The wizard continues to search for USBXpress Device e During this process a prompt box may pop up and click Continue Anyway...

Страница 41: ...n INSTALL USBXpress Device MANUALLY z These steps should be followed when USBXpress Device cannot be automatically installed successfully a Right click the shortcut My Computer on the desktop and ente...

Страница 42: ...of the analysis to generate analytical reports Data receiving process is as the following steps 1 Software initiation Firstly double click shortcut of the software icon on the desktop the system will...

Страница 43: ...the Connect button will be grayed And other originally grayed unavailable buttons will be available now After online with the computer see online between recorder and computer recorder will automatica...

Страница 44: ...record start In fact in the process of data transmission the system can automatically identify the actual end time of record Recorder will automatically stop after data transmission is finished So yo...

Страница 45: ...4 Open data file dialog box First of all through the drop down menu of the window select the menu of sleep data file you can also click the mouse to select the menu Selected menu the data file lists...

Страница 46: ...em main window then chooses some data file demonstrated then may switch to the different data file window Chooses in the toolbar or select cancel in the File option in the menu may close the current w...

Страница 47: ...shorten the time axis which is favorable to view the trend change of recorded parameters according to the different time resolution The cursor on the trend chart can be moved around through the button...

Страница 48: ...be deleted stay the cursor at the exact point and click the button directly to erase them Event location Click on the chart to make the cursor stay anywhere it is wanted to be Also clicking on the ic...

Страница 49: ...ts Other operations in the current window involve editing analysis result inserting changing and deleting all events Figure 6 Waveform display The following operations may carry out in this page 1 Pri...

Страница 50: ...evious or next one minute upon each clicking on the icons 3 The bar is a time vernier which could be dragged back and forth or clicked on to show a waveform in varying time ranges 4 Moving the mouse c...

Страница 51: ...lock on the waveform when a menu pops up demonstrating the length of this block Select a corresponding event type to generate such an event Mark function When the operator feels particularly intereste...

Страница 52: ...tion to revise or delete an event When the mouse arrow becomes draggable at the beginning or end of an event block drag it back and forth to shorten or lengthen the time of the respiration event Also...

Страница 53: ...spiration Event 4 Report All analysis results are produced into summary report respiration analysis report SpO2 analysis report after statistical analysis If the data recording has contained the CPAP...

Страница 54: ...the drop down menu All the above reports can be printed out Figure 10 Diagnosis result 10 2 4 Print preview and output When the waveform is in need of printing the Raw Data page should be activated to...

Страница 55: ...button in the toolbar or Print from File menu will show up the window for preview options The print work will be done after selecting the intended reports and pressing the button Click to change the p...

Страница 56: ...lymate YH 1000C Polysomnograph User Manual V3 1 52 Figure 13 Alter Report Title 10 2 6 Exit program To back to Windows desktop just click File in the menu and click Exit or click on the top right corn...