Auto CPAP / CPAP System User Manual V1.0
37 / 51
18. Cleaning and Disinfection
WARNINGS!
• Regular cleaning of the device and its accessories is very important for the prevention of
respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
• Use mild soap that is nontoxic to humans.
• Follow the manufacturer's instructions on cleaning the mask and tubing and on determining
the frequency of cleaning.
• Before cleaning, check whether the device has been disconnected from the power supply,
whether the power cord has been unplugged, and whether the water chamber of the device
has cooled down. Make sure the plate has cooled down to room temperature, so you do not
get burned.
• Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorized service agent.
CAUTIONS!
• Overheating of the materials could lead to early fatigue of these materials.
• Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean the device
and its accessories. Liquid soap containing moisturizing agents or antimicrobials should not
be used either. These solutions may harden cleaned materials or reduce their lifespan.
• Do not clean or dry the device and its accessories when the temperature is higher than
80°C (176°F). High temperatures could reduce product life.
• Do not immerse the device in any fluids.
18.1 Cleaning the Mask and Headgear
For details, refer to the cleaning instructions in the user manual for the mask.
18.2 Cleaning the
SpO
2
Kit
Wipe the surface of the SpO
2
Kit with a clean, soft, and slightly damp cloth.
CAUTIONS!
• It is recommended to clean the SpO
2
Kit once a week.