Auto CPAP / CPAP System User Manual V1.0
17 / 51
- Use only masks with vent holes. Do not block or try to seal the vent holes in the exhalation
port.
11.5 Using Oxygen with the Device
Oxygen may be added at the mask connection. Please observe the instructions listed below
when using oxygen with the device.
WARNINGS!
• Connect the oxygen tube to the oxygen inlet of the mask.
• The oxygen supply must comply with the local regulations for medical oxygen.
• Turn on the device before turning on the oxygen. Turn off the oxygen before turning off the
device. Explanation of Warning: When the device is turned off, but the oxygen flow still exists,
oxygen may accumulate within the device's enclosure and pose a fire hazard. Turning off the
oxygen before turning off the device will prevent oxygen accumulation in the device and
reduce the risk of fire. This warning applies to CPAP devices.
• Oxygen supports combustion. Keep the device and the oxygen container away from heat,
open flames, any oily substances, or other sources of ignition. DO NOT smoke in the area
near G3 A20 / G3 C20 or the oxygen container.
• Sources of oxygen should be located more than 1 m from the device.
• When using oxygen with this system, a Pressure Valve must be placed in-line with the
patient circuit between the device and the oxygen source. The pressure valve helps prevent
the backflow of oxygen from the patient circuit into the device when the unit is off. Failure to
use the pressure valve could result in a fire hazard.
• Do not connect the device to an unregulated or high pressure oxygen source. The pressure
of oxygen source does not exceed the work pressure of the device.
11.6 Inserting the SD Card
(
Only for the device that
equipped with SD card)
Insert the SD card into the SD Card Slot, as shown in Fig. 11-9.
SD Card Slot
Fig. 11-9
If the SD card is inserted correctly, a symbol
indicating correct insertion will appear in
the Main Interface on the screen of the device.
If the SD card is inserted incorrectly, a symbol
indicating incorrect insertion will appear in
the Main Interface on the screen of the device.