BMB MEDICAL
–
I-MOVE
–
2VTRS062022
15
2.4
PRODUCT
POWER
SUPPLY
Power grid connection conditions and electromagnetic compatibility:
This medical device’s
power cord may represent an obstacle that could cause a fall. Be
mindful of its presence when moving around the appliance, when the latter is connected to
the mains.
Position the product such that it can be easily unplugged.
Unplug the product before moving it.
Do not leave the cables in contact with the floor or the casters,
Do not knot the cables,
Periodically check the condition of the cables.
Mains connection:
100~240V depending on the country.
Frequency:
50 /60 Hz
Rated current:
3.3 A
Protection rating / degree of electric shock
protection:
Class 2 appliance
Power input:
525 W
Intermittent operating mode:
5% i.e., 1 minute of continuous operation followed
by 19 minutes of rest.