6.0 Troubleshooting guide
Trouble
Correction
Conductivity pen
gives low readings
Low readings usually mean the probe is contaminated. Clean the
probe and retest in a known solution. Ensure unscented cleaner is
used eg. Bluelab Conductivity Probe Cleaner, Jif, Liquid Vim, Soft
Scrub, Cif cream or Viss.
Conductivity pen
gives high readings
Calibrate pen in a known standard solution.
Check the table in section 4.0 for what solution to use for your
selected conductivity unit.
Screen does not
turn on
Replace unit.
7.0 Technical specifications
Measurement Range
0.0 - 10.0 EC
0 - 5000 ppm (500ppm / TDS)
0 - 7000 ppm (700 ppm)
0 - 50 ˚C / 32 - 122 ˚F
Resolution
0.1 EC, 10 ppm (700), 10 ppm (500)
1 °C / 1 °F
Accuracy at 25 ˚C / 77 ˚F
± 0.1 EC, ± 50 ppm (500 ppm), ± 70 ppm (700 ppm)
± 1 ˚C / ± 2˚F
Temperature compensation
Automatic
Operating temperature
0 - 50 °C / 32 - 122 °F
Calibration
Factory calibrated / manual calibration optional
Units
EC, 700 ppm, 500 ppm, °C, °F
Power source
1 x AAA alkaline battery
English
6