1. SAFETY AND USER INFORMATION 

Congratulations to the ownership of this high quality product. 
A long life expectancy is assured if used in the described 
manner and correct application. As with all technical 
products, a hazard to health or equipment can exist if 
improperly used, the unauthorized removal of necessary 
covers, incorrect installation or incorrect operation is 
present. Follow these instructions and adhere to the generally 
accepted rules of technology. Installation and setting-up 
should only be carried out by qualified service personnel ( IEC 
364 / VDE 0105).
 

1.1 PACKAGING 

Carefully check the equipment immediately after receipt, for 
transport damage, deformation, and loose parts. Any damage 
should be reported without delay to the transport carrier, 
even then when no apparent damage to external package s 
visible.

1.2 STORAGE

permitted storage
temperature: -25…+85 °C
permitted humidity: 30…80 % relative humidity
Residual condensation is not permitted during operation

1.3 INSTALLATION AND OPERATION

This equipment is to be protected against improper use. 
Components are not to be bent or isolation spacing to be 
changed, when transported or handled. The contact with 
electrical components and terminals is to be avoided.

Always disconnect the equipment from the mains supply 
before installation or wiring is started!

1.4 MAINTENANCE AND SERVICING

Electrical equipment generally requires no special 
maintenance, are however (depending on the protection index) 
to be protected against dust build-up, moisture, radiation 
and aggressive chemicals. Servicing is only permitted under 
the terms and conditions of these operating instructions. 
Nevertheless should a failure occur, please return the 
equipment to us for repair giving the following information: 
Type of fault, accompanying symptoms (operation conditions), 
your own speculation as to the cause of the failure, previous 
unusual conditions etc.

To maintain reliable operation of the laboratory power supply, 
it is advisable after approximately 8000h operation or two 
years, to inspect and service the built in variable transformer.

a) Check for free movement of the carbon brushes/ 
contact roller. It is important that there is adequate 
contact pressure between the carbon brushes/roller and 
the area of the winding/ contactor.

b) Jammed, worn or damaged brushes/rollers should be 
exchanged immediately, otherwise damage or destruction 
to the winding or contactor could result.

c) The contact area of the winding should then with a hard 
brush, be cleaned of all dust and dirt particles. An abrasive 
cleaning powder may be used, but after a successful 
cleaning process, any residual powder must be thoroughly 
removed.

d) The contactors should be checked for signs of wear and 
accordingly exchanged.

e) With abnormal operating conditions1)  according to 
VDE 0552 G the duration between servicing is reduced to 
approximately 1500 – 2000 operation hours.

1) Abnormal operation is when during a 24 hour 
operation period no adjustment of the variable 
transformer is made.

       

1.5 DISPOSAL

Please observe the current regulations and dispose according 
to type of material i. e. as electronic scrap (printed circuit 
boards), plastics (housing), steel, copper etc.

1.6 AMENDMENTS

We have produced this documentation with the utmost care, 
however no guarantee for completeness or correctness can 
be given. The adoption of this information in an application 
must be individually checked. The technical details describe 
the product features without guaranteeing these. This 
product is subject to changes that serve the technical 
advancement.

2. GENERAL

The Laboratory AC power supply type BR 1000 gives a 
continuously adjustable output voltage between 0 and 250 
volts ac. The output circuit is isolated from the mains supply. 
The incorporated isolation transformer has a screen winding 
which prevents mains interference being transferred to the 
secondary. Through a circuit breaker, the input and output 
circuits are safeguarded. The mains switch is illuminated. 
The housing has two inset hand grips which makes it easy to 
stack and is therefore room saving. The output socket is not 
earthed. Due to the protection index (IP 20), the use of this 
equipment is only permitted in dry areas.

BR 1000

INSTRUCTION MANUAL
LABORATORY AC POWER SUPPLY

B

R 1000  2013•

09

Subject to change.