10
When fully charged the device’s battery is intended to provide up to 3 days of normal use
(conditional on the type of use). We recommend that the knee be fully charged every night.
Switching off the device when not in use will also prolong the period of use
Note… The device will take 30 seconds to reset after it is switched back
ON
[
I
] ; refer to Section 6
Battery Charging (continued) for the reset sequence.
Battery Charge Indicator
LED
Beeps
Battery Condition
Green
3 flashes
--
fully charged
Green
2 flashes
--
good
Green
1 flash
--
ok
Yellow
slow flash
8 x Short beeps
low charge
Red
slow flash
5 x Long beeps
no charge
Device
LED
6.1
Magnetic Charger Connection
The charging connector can be connected to the device through the charging terminal located
at the posterior of the knee.
The connector and the terminal magnetically lock together and correctly self-align on
connection (Figure 1).
Please check for any misalignment between mating surfaces as
this may result in no charging. This can happen especially when
the charging setup puts strain on the charging cable (Figure 3).
Before charging ensure charging surfaces are clean, dry and free
of any debris.
Do not allow any ferrous material to attach itself to the charging
surfaces.
Keep the magnetic surfaces away from materials that may be
adversely affected by their proximity to strong magnetic fields
(e.g. watches, pacemakers, bank cards or magnetically stored
media).
Prevent build up of debris on the charging surfaces as this may
prevent charging or damage the charger. Disconnect the charger
from the mains socket and remove any debris before reconnecting
the charger to the device.
1.
2.
3.
Note… The connector and the terminal are not polarized so
orientation is not critical (Figure 2).
6.2
Battery Status
To check the battery condition press and
hold
the [
-
] button on the knee, one short beep will be
heard and the LED will indicate the battery condition.
Device
938354V2APK1/1-0321