5.1
Battery Level Indication
LED
Battery Charge
3 flashes
Maximum
2 flashes
High
1 flash
Low
5.2
Device Status Indication
The same LED on the left hand side of the device is also used to indicate device status:
1.
Single Flash
When the charger is disconnected it will flash once to indicate correct start up.
2.
Steady ON
If after the charger is disconnected it stays on it indicates that an error was detected by your
device.
3.
Flashing rapidly
If it flashes rapidly, it indicates that
Bluetooth®
is active. This capability is to be used only by
the clinician. The flashing can be ignored and will cease after a few minutes. Avoid holding
the ankle horizontally for prolonged periods of time to prevent activating
Bluetooth®
.
See Appendix 1 for a full summary of all possible LED indications of the device and the charger
pod.
To check the battery charge level,
tilt the foot down vertically and hold
stationary for 4 seconds. The Status
Indicator LED on the left hand side of
the device will flash blue, 1-3 times
(see table) and then repeat to indicate
the charge status. Tilt the foot again to
repeat the battery status check.
Note: This feature can be switched off.
Note: After charging the device fully
the Status Indicator LED may indicate
2 or 3 flashes. This will depend on the
charging time, the age and condition of
the battery.