The device and its charger must not be used by users of
pacemakers or implantable cardioverter-defibrillators (ICDs).
Contact your service provider for advice if the device is to be
used with any other electronic medical device.
2. Only charge the batteries in the device using the charger
supplied with the product.
Do not use this charger for charging any other device.
3. The battery charger can only be used with an electrical outlet that is rated
between 100 V and 240 V AC, 50/60 Hz.
4. Never connect the charger to the device while still wearing the limb.
5. During charging no power is delivered to the ankle control system.
6. The ankle and the charger may become warm during charging but should never be
too hot to touch. There should be no abnormal smell. If this is the case immediately
unplug from electrical supply and contact your service provider.
7. Do not refit the limb until the charger has been disconnected.
8. Always recharge the device before use if it has not been used for an extended
period of time.
9. The inductive battery charger kit is not waterproof. Do not charge in atmospheres
containing liquids and/or powders.
10. Do not exceed the charging temperature limitations set out in this manual.
11. When charging an empty battery the charger may terminate prematurely (after a
few minutes). If this happens, reconnect the charger.
If this problem recurs contact your service provider.
12. Do not leave the charger plugged to electrical supply after use.
Charging Error
Charging error is always indicated by a non-responding Charging Indicator LED on the right
hand side of the device (no initial flash or charging state indication).
LED
Error
Corrective Action
Slowly Pulsating No Error/Normal charging Not needed
Steady Blue
Light
Charger operational but
not mated correctly to the
charging surface of the
device.
Try reattaching and check if
anything prevents perfect
mating of charging surfaces.
Remove any intruding
material.
No Light
The charger is not
powered.
Check that charger is
connected to the mains.
Rapid Pulsating Possible overheating of
the charger.
Charging may resume
when normal charging
temperature is reached.
If recurring contact your
service provider.
Further information is provided by the charger pod LED (blue):
Warnings
Charger
Pod LED