15
|
G7 Dock Technical User Manual
©2022 Blackline Safety Corp
2022-09-26
Rev. 11
4.
Press G7’s arrow buttons to navigate to Calibration on the LCD screen menu and press
G7’s OK button to select. G7 Dock will perform the calibration.
5.
G7 will indicate whether the calibration has passed or failed and when it is ready to be
removed from G7 Dock.
4.4
G7 DOCK UPDATE
Updates to your Dock’s settings can be made in your G7 Dock’s configuration profile, from
the Docks page on Blackline Live. The activation code on the back of Dock can be used to
identify it from the list. If you are having difficulties accessing this page or finding G7 Dock in
Blackline Live, please contact our
NOTE:
G7 Dock does not need to be connected to the Blackline Safety Network to charge a
G7 device or complete calibrations and bump tests, but the G7 device does need to be
connected to the network to update G7 Dock configuration settings.
4.5
MULTI-DOCK SETUP
WARNING:
Do not use a multi-dock setup if you are calibrating with Chlorine (Cl
2
). Calibration
with Cl
2
requires that the tube length be kept as short as possible, which is not supported by a
multi-dock set-up.
In situations where multiple safety monitoring devices require calibration and bump testing,
docks can be joined together to perform these functions across all connected devices.
Multiple G7 Docks can be connected to a single-gas cylinder by using tubes, check valves, and
T-connectors. One gas cylinder can support five G7 Docks at one time. Blackline offers a
multi-dock setup kit to help with multi-dock setup. To order, please contact our Customer
Care team or your distributor.
Required parts and tools:
•
G7c or G7x personal safety monitoring devices (single-gas or multi-gas cartridge)
•
G7 Docks for each device
•
Tube (1/8” inner diameter)
•
Nylon or polyethylene T-connectors
•
Nylon check valves
•
Gas cylinder with specific single-gas or multi-gas mixture and a demand flow regulator
•
120/240 VAC power source for each dock