E13C2 User Guide
(16-01484-EN-01)
Cleaning after Puncture and Biopsy
Use a suitable brush to make sure that biological material and gel are removed from
all channels and grooves. See
Care and Cleaning
for cleaning instructions.
Disposal
When the transducer is scrapped at the end of its life, national rules for the relevant
material in each individual land must be followed. Within the EU, when you discard
the transducer, you must send it to appropriate facilities for recovery and recycling.
WARNING
Reproc-w3
Immediately after use, you must pre-clean the device until visually clean (including
device lumens if existing). Conduct the thorough cleaning process as soon as
possible after use in order to prevent bioburden drying on the surface. Dried
bioburden can lead to inefficient cleaning, disinfection and sterilization, causing a
risk of cross-contamination.
If pre- and thorough cleaning cannot be done immediately, keep the device moist until
cleaning.
WARNING
D-w1
For disposal of contaminated items such as transducer covers or needle guides or other
disposable items, follow disposal control policies established for your office, department, or
hospital.