28
*RST Value
MAXimum
Example
CURR:PROT 3
Query Syntax
[SOURce:]CURRent:PROTection[:LEVel]?
Query Response
NR2
7.14 [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
This command is used to set the current setting in CC mode.
Command Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters
MIN MAX- MINimum- MAXimum
Unit
A
*RST Value
MINimum
Example
CURR 5
Query Syntax
[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?
Query Response
<NR2>
7.15 [SOURce:]POWer:PROTection[:LEVel] <NRf+>
This command sets the power protection level. If the input power exceeds the power protection level for the time specified
by POW:PROT:DEL, the input is turned off.
NOTE:
Use POW:PROT:DEL to prevent momentary power limit conditions caused by programmed changes from tripping
the over power protection.
Command Syntax
[SOURce:]POWer:PROTection[:LEVel] <NRf+>
Parameters
MIN MAX - MINimum - MAXimum
Unit
W
*RST Value
MAXimum(high range)
Example
POW:PROT 100
Query Syntax
[SOURce:]POWer:PROTection[:LEVel]?
Query Response
<NR2>
7.16 [SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRf+>
This command is used to set the power setting in CP mode.
Command Syntax
[SOURce:]POWer[:LEVel][:IMMediate][:AMPLitude] <NRf+>
Parameters
MIN MAX- MINimum- MAXimum