12
INT MEMORY ERROR
LOADING DEFAULTS
Next screen directs the user to the menu to select the sub-menu. The battery is auto-sensed & the defaults are loading accordingly.
After memory test, the open circuit voltage is displayed on the first line of LCD, the required Ah value can be selected using up & down keys.
OPEN CKT: 12.53V
SELECT Ah: 028
On selecting the required Ah value, press Select Ah & then press the TEST key, the measurements will begin & following message is displayed.
Backlight will switch off during the test.
TEST INPROGRESS
12.53V 028AH
The results will be displayed as follows.
12.37V 61.87%CAP
IR: 39.2 mOhm
The first field indicates the battery terminal voltage when battery was loaded during the test; the second field shows the balance capacity in
percentage. IR indicates the internal resistance in milli-ohms.
After the test, follows a cooling period of approximately 15 seconds. If user tries to perform a test during this period, the following message is
displayed.
TO BEGIN NEXT
SESSION, WAIT….
In case the open circuit voltage of BUT exceeds 14.8V, the following screen will be continuously flashed & you will not be able to enter the test
mode.
X HIGH VOLTAGE X
REMOTE BATTERY