23
1
The offset voltage has three ranges as follows:
Output Voltage
Range
Constraints of
Amp Offset
1.01 volt to
10.00 volts
(Vp-p)/2 + |offset| <= 5
volts
0.101 volt to 1
volt
(Vp-p)/2 + |offset| <=
0.5 volts
0.010 volt to
0.100 volt
(Vp-p)/2 + |offset| <=
0.05 volts
F4:SYM
– (Symmetry)
When the Square waveform is
selected, the SYMMETRY (Duty cycle) parameter is
available. Change the duty cycle (square) by using the
cursor keys, rotary dial or numerical keys. If a certain
setting cannot be produced, the waveform generator will
display a warning message.
Square
Menu