15
1
2.4 Instrument Mounting
The 4007B and 4013B Function Generators are intended
for bench use. The instrument includes a front feet tilt
mechanism for optimum panel viewing angle. The
instrument does not require special cooling when
operated within conventional temperature limits. It may
be installed in a closed rack or test station if proper air
flow can assure removing about 15 W of power
dissipation.
2.5 Power Requirements
The 4007B and 4013B can be operated from any source
of 90V to 264V AC, frequency from 48Hz to 66Hz. The
maximum power consumption is 25 VA. Use a slow blow
fuse of 1A, UL/CSA approved as indicated on the rear
panel of the instrument.
The instrument power fuse is located in the AC input
plug. To access the fuse, first disconnect the power cord
and then remove the fuse cartridge.
2.6 Fuse Replacement
There is a 1A, 250V rated slow blow fuse at the AC input.
Should the fuse ever get blown, follow the steps below to
replace: