SHOCK HAZARD:
Getting Started
Before connecting and powering up the instrument, review the instructions in this section.
2.1
Input Power Requirements
The oscilloscope has a universal AC input that accepts line voltage and frequency input within:
Line
Voltage Range
100-120 V
100-240 V
Frequency
400 Hz
50/60 Hz
Power
80W Max
Table 2.1
Before connecting to an AC outlet or external power source, be sure that the power switch is in the OFF position and
verify that the AC power cord, including the extension line, is compatible with the rated voltage/current and that there
is sufficient circuit capacity for the power supply. Once verified, connect the cable firmly.
The included AC power cord is safety certified for this instrument operating in rated range. To
change a cable or add an extension cable, be sure that it can meet the required power
ratings for this instrument. Any misuse with wrong or unsafe cables will void the warranty.
The power cord provides a chassis ground through a third conductor. Verify that your power
outlet is of the three-conductor type with the correct pin connected to earth ground.