2052 User Guide
(BB1804-C)
25
3D Imaging
To acquire a 3D dataset, you must have the 3D option installed on your system: this
includes the necessary software and hardware as well as a 3D license.
To acquire a 3D dataset:
•
See the system user guide for instructions.
NOTE:
The maximum length of a dataset acquired using this transducer is 60mm. The
resolution of the dataset is never worse than 0.5mm (i.e. 2 images per 1mm of
movement). The default resolution is 0.25mm.
Transducer Support Collar UA0679
Figure 18. The support collar on the transducer without and with a fixation clamp.
Slide the collar over the handle of the transducer and clamp it in place.
Then with the variable friction support arm, you can either:
•
Clamp the instrument fixation clamp to the support collar.
or
•
Remove the fixation clamp from a support arm and screw the support collar
directly to the arm, using the threaded hole at the end of the steel ball joint piece.
Disinfection and Sterilization
The transducer support collar can be autoclaved.
Disposal
When the transducer is scrapped at the end of its life, national rules for the relevant
material in each individual land must be followed. Within the EU, when you discard
the transducer, you must send it to appropriate facilities for recovery and recycling.
See the applicable system user guide for further details.
Caution:
It is strongly recommended that you use the transducer support collar to
reinforce the transducer handle when you use a fixation device (such as the variable
friction support arm UA0553) to hold the transducer in place.
WARNING
For contaminated disposals such as transducer covers or needle guides, follow
disposal control policies established for your office, department, or hospital.