10
January 2014
2052 User Guide
(BB1804-C)
Selecting the Crystal and Imaging Frequency
To select the other crystal or change the imaging frequency, please refer to your
system user guide.
The 2052 Crystal Position Imaging With Pro Focus UltraView 2202
This section
only
applies to Anorectal Transducer Type 2052 when used with the Pro
Focus UltraView 2202 system.
The 2052 Crystal Position Displayed on the Pro Focus UltraView 2202
Screen
The system tracks the crystal position and displays it in cm on the left side of the
screen next to
Crystal Pos
.
Figure 5. Position of the 2052 transducer crystal.
Default
position
The crystal position is 0 when the crystal is at the tip of the transducer (farthest into
the patient). This is the default position.
The following actions cause the crystal to move to the default position:
•
Connecting the transducer to the system.
•
Selecting a new preset.
•
Pressing the
Forward
button until the system beeps.
•
Entering a new patient ID on the screen.
Other
positions
All other crystal positions are expressed as negative numbers (see Fig. 5 for an
example). -6.0 (cm) is the crystal position closest to the transducer handle.
Changing the 2052 Crystal Position on the Pro Focus UltraView 2202
You can use the
Forward
and
Backward
buttons on the transducer handle to move
the crystal.
You can also use the on-screen control to change the crystal position.
To move the crystal in the 2052 transducer:
Transducer
control
button
•
Press one of the control buttons on the upper side of the transducer handle (see
Fig. 4) until the crystal is where you want it.
anatomic (body)
position (in cm)
crystal position (in cm)
crystal position slider
insertion depth (in cm)
destination position
Содержание Type 2052
Страница 1: ...English BB1804 C January 2014 For Professional Users Only User Guide Type 2052 Anorectal 3D Transducer...
Страница 4: ...4...
Страница 26: ...26 January 2014 2052User Guide BB1804 C...
Страница 27: ......
Страница 28: ......