12

 November 

2018

N13C5 User Guide

 (16-01270-EN-06)

Puncture

lines indicate

expected

needle path

The puncture line on the scan image is an indication of the expected needle path (see 
Figure 7). The distance from the guide channel entrance of the puncture attachment 
to the first dot on the scan image puncture line is approximately 7.5 mm (0.3 in) for 
25°, 12 mm (0.5 in) for 45°, and 19 mm (0.7 in) for 65°. The distance between the 
dots is 5 mm (0.2 in).

Assembling and Mounting the Needle Guide

To assemble and mount a needle guide on N13C5:

1

Apply a small amount of imaging gel to the tip of the transducer and carefully 
cover the transducer with a sterile cover.

2

Select the required needle guide insert by breaking it off the palette (see Figure 
2).

3

Find the wedge-shaped channel in the needle 
guide, and slide the insert into the channel in 
the needle guide until it clicks into place and 
aligns with the needle guide.

4

Hold the assembled needle guide with the bracket pointing down.

5

Align the two grooves on the transducer with the two ridges on the needle guide.

6

Carefully smooth and stretch the 
transducer cover along the face of the array 
and apply pressure until the needle guide 
clicks into place over the end of the 
transducer. 
Figure 4 below shows the needle guide 

mounted on the transducer.

Содержание N13C5

Страница 1: ...For Professional Users Only 16 01270 EN 06 November 2018 N13C5 Transducer User Guide...

Страница 2: ...ufacture of a product please contact your BK Medical representative or write to us at the email address above including the product s serial number SN number BK Medical Customer Satisfaction Input fro...

Страница 3: ...er Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Neurosurgical Imaging with N13C5 9 Puncture and Biopsy Facilities 10 Single Use Needle Guide UA1345 for N13C5 10 Assembling and...

Страница 4: ...4...

Страница 5: ...aniotomy Patient Population The patient population is neonates2 infants children adolescents and adults Physicians only Caution Rx c1 United States law restricts this device to sale by or on the order...

Страница 6: ...ns ortheimageisseverelydistortedordegraded oryou suspect in any way that the system is not functioning correctly Remove all transducers from contact with the patient Turn off the system Unplug the sys...

Страница 7: ...nd repair of BK electromedical equipment must be carried out only by the manufacturer or its authorized representatives BK Medical reserves the right to disclaim all responsibility including but not l...

Страница 8: ...da use only licensed transducer covers In Europe transducer covers must be CE marked WARNING GS w4a It is essential for the patient s safety that only the correct equipment is used Do not use other ma...

Страница 9: ...lowing the instructions for the puncture attachment Verify that the cover has not been damaged in the process If it has repeat the procedure with a new transducer cover Using the Transducer Control Bu...

Страница 10: ...e humidity of 0 to 80 WARNING TC w5 If the transducer cover is damaged during interventional procedures follow the policies of the hospital or clinic for treatment of the patient under such circumstan...

Страница 11: ...ith free angle inserts The 3 angle inserts each contain three channels These channels allow the needle to be positioned at 25 45 or 65 to the image axis of N13C5 The free angle inserts are slotted to...

Страница 12: ...unt of imaging gel to the tip of the transducer and carefully cover the transducer with a sterile cover 2 Select the required needle guide insert by breaking it off the palette see Figure 2 3 Find the...

Страница 13: ...a LEYLA arm NOTE Ensure that the transducer cable is secured so that it cannot move and pull the transducer out of position WARNING P w9 Ensure that the channel bracket and needle guide are correctly...

Страница 14: ...ith a new transducer cover before reattaching the needle guide WARNING Exam w5 Be careful not to lose any parts of the needle guide inside the patient WARNING P w1 Before you start imaging verify that...

Страница 15: ...e use needle guide UA1345 when mounted on N13C5 transducer You must take extra care when taking a free angle biopsy because the expected needle path is not shown on the screen image Cleaning after Pun...

Страница 16: ...is scrapped at the end of its life national rules for the relevant material in each individual land must be followed Within the EU when you discard the transducer you must send it to appropriate faci...

Страница 17: ......

Страница 18: ......

Отзывы: