Cleaning after Puncture and Biopsy

Use a suitable brush to make sure that biological material and gel are removed from 

all channels and grooves. See 

Care and Cleaning

 for cleaning instructions.

3D Imaging

3D imaging with the E14C4t is possible:

Freehand – where the transducer is combined with the appropriate 3D system 

software.

Disposal

When the transducer is scrapped at the end of its life, national rules for the relevant 

material in each individual land must be followed. Within the EU, when you discard 

the transducer, you must send it to appropriate facilities for recovery and recycling.

WARNING

Reproc-w3

Immediately after use, you must pre-clean the device until visually clean (including 

device lumens if existing). Conduct the thorough cleaning process as soon as 

possible after use in order to prevent bioburden drying on the surface. Dried 

bioburden can lead to inefficient cleaning, disinfection and sterilization, causing a 

risk of cross-contamination.

WARNING

D-w1

For disposal of contaminated items such as transducer covers or needle guides or other 

disposable items, follow disposal control policies established for your office, department or 

hospital. 

Содержание E14C4t

Страница 1: ...For Professional Users Only 16 01265 EN 07 June 2019 E14C4t Transducer User Guide...

Страница 2: ...a BK Medical transducer contains information about the date of manufacture BK Medical Customer Satisfaction Input from our customers helps us improve our products and services Your opinions are impor...

Страница 3: ...Buttons 9 Changing Orientation 9 Imaging with E14C4t 9 Simultaneous Biplane Imaging 9 Transverse or Sagittal Imaging 10 Endfire Imaging 10 Imaging Without Puncture or Biopsy 10 Adjusting Image Area a...

Страница 4: ...4...

Страница 5: ...ing and endfire imaging in a single transducer E14C4t is designed for transrectal and transvaginal procedures It is also suitable for elastography and contrast imaging Patient Population The patient p...

Страница 6: ...Information Product specifications acoustic output data and data about EMC electromagnetic compatibility for this transducer can be found in the Product Data Sheet and the Technical Data BZ2100 that...

Страница 7: ...This can damage the cable and lead to degradation and or corruption of the ultrasound image WARNING AO w1 To avoid tissue damage always keep the exposure level the acoustic output level and the expos...

Страница 8: ...e CE marked WARNING T w5 To prevent electrical shock and damage to the transducer the connector pins in the transducer plug must always be completely dry before you connect to a system WARNING GS w4a...

Страница 9: ...ton enables the sagittal or endfire array A long press on this button also activates endfire imaging A second button enables the transverse array Changing Orientation To change the orientation of the...

Страница 10: ...t enable split screen imaging as there is no defined isocenter in this context When you press the system s Split key tap DUAL two images appear on the screen The endfire image appears and either a tra...

Страница 11: ...in both the transverse and sagittal planes Endfire imaging in the sagittal plane Dual combines both the simultaneous biplane and endfire imaging in one biopsy guide All the biopsy guides for transrect...

Страница 12: ...A1326 are used for simultaneous biplane imaging Figure 5 Reusable biopsy guide UA1326 The puncture line for UA1322 S UA1322 S14 and UA1326 on E14C4t is shown in Figure 6 WARNING Sterile w1 Single use...

Страница 13: ...guide UA1327 The puncture line for UA1323 S and UA1327 on E14C4t is shown in Figure 8 The needle guide is parallel to the centerline of the transducer Figure 8 Illustration of the puncture line for b...

Страница 14: ...ile imaging gel over the transducer 3 Slide the needle guide into the biopsy channel bracket 4 Insert the assembled needle guide and biopsy channel bracket into the open channel on the transducer A sm...

Страница 15: ...he needle guide can be adjusted 70 mm lengthwise with respect to the mounting ring using the adjustment screw Figure 12 Puncture attachment UA1324 Mounting the Transperineal Puncture Guide To mount th...

Страница 16: ...User Guide 16 01265 EN 07 Figure 13 Puncture attachment UA1324 mounted on E14C4t The puncture lines for UA1324 on E14C4t are shown in Figure 14 Figure 14 Illustration of the puncture line for punctur...

Страница 17: ...instructions WARNING P w1 Before you startimaging verifythatthetypenumberornameofthetransducerandthetype numberordescriptionof the needleguideyouareusing matchthenumberdisplayedonthe monitor Also make...

Страница 18: ...must be followed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling WARNING Reproc w3 Immediately after use you mustpre cleanthe devic...

Страница 19: ......

Страница 20: ......

Отзывы: