Sterile Biopsy Guides (Dark Green)
The sterile single-use biopsy guides (UA1322-S, UA1322-S14, UA1323-S and
UA1329-S) come assembled in peel packs. Contents are only sterile if the package
is intact.
Recommended needle length is 20 cm (7.8 inches) for anesthetic procedures and 25
cm (9.8 inches) for biopsy procedures.
The sterile-packed biopsy guides must be stored at a temperature range from +5°C
(+41°F) to +25°C (+77°F) and at a storage humidity of 0% to 80%.
Simultaneous Biplane
The sterile single-use biopsy guide UA1322-S, UA1322-S14 and reusable biopsy
guide UA1326 are used for simultaneous biplane imaging.
Figure 5. Reusable biopsy guide UA1326.
The puncture line for UA1322-S, UA1322-S14 and UA1326 on E14C4t is shown in
WARNING
Sterile-w1
Single-use components are packaged sterile and are intended for single-use only.
Do not use if:
•
integrity of packaging is violated
•
expiration date has passed
•
package label is missing
WARNING
Sterile-w2
Sterile-packed components must be stored in a safe environment and kept out of direct
sunlight. Large temperature changes during storage may cause condensation and violate
the integrity of the packaging.
WARNING
D-w1
For disposal of contaminated items such as transducer covers or needle guides or other
disposable items, follow disposal control policies established for your office, department or
hospital.