P a g e
|
9
XXPRESS® NGX USER MANUAL
– XP-5001-01-5
4
Technical Specification
Model Type
Model Name
Model Numbers
xxpress
®
NGx Thermal Cycler
XP-1003-CC1 (220V-240V)
XP-1004-CC1 (100V-120V)
Physical Unit Properties
Dimensions (W x D x H)
Weight
300 mm x 520 mm x 315 mm
35 kg
Mains Supply
Power Cord Rating
Inlet Module Type
Supply Voltage
Supply Frequency Range
Power Consumption
Fuse Ratings and Sizes
Mains Supply capability
IEC C13, 3-Core, 10A min
Only use an appropriately rated and
approved supply cordset in accordance
with the regulations of the country in
which the equipment is put into service
IEC C14, DPST, Single Fuse
100V-120V AC
(XP-1002-01)
220V-240V AC
(XP-1001-01)
50Hz/60Hz
1000W max
T8A/250V HRC 20x5mm
The power for this unit must come from
a stable mains supply. Do not power
from inverters, frequency convertors, or
generators.
(See Section 5 for further details)
Operating Environment
Temperature Range
Relative Humidity Range
Maximum Operating Altitude
+18 to +30
C
(Ambient temperatures above 25
C
can reduce the rate of cooling)
20% to 80% non-condensing
2000m above sea-level