13
Set Hi SETTING UPPER THRESHOLD
By pressing
ZERO
the display shows “000.000kg”, with the first digit blinking, and it’s therefore possible to configure
the upper threshold:
-
Use the
F
key to clear the value, the
TARE
key to increase the value of the digit, the
U
key to go on the previous
digit and the
SMPL
key to go on the following digit.
-
Confirm with
ZERO
the entered value or press
PRINT/M+
to exit without confirming.
7.3 “F1 toL” – CLEARING TOTAL
toCLr CLEARING TOTAL
By pressing
ZERO
it’s possible to clear the accumulated total weight.
Confirm with
ZERO
or press
PRINT/M+
to exit without confirming.
to P-C
Not used in this application.
to Prt
Not used in this application.
7.4 “F2 Unt” – CONFIGURATION OF SECONDARY UNITS OF MEASURE
By pressing
ZERO
it’s possible to choose the units of measure for the conversion function (see section 6.8), enabling
(on) or disabling (oFF) the suggested secondary units of measure:
1) The display proposes the first secondary unit of measure to be enabled or disabled.
2) Press
TARE
to change the options and confirm with
ZERO
,
or press
PRINT/M+
to exit without confirming and
pass to the following unit.
3) The display proposes the 2nd secondary unit to be enabled or disabled.
4) Proceed as described in point 2 up until the last proposed unit of measure.
7.5 “F3 oFF” – ENABLING DISPLAY BACKLIGHTING, AND ACOUSTIC ALARM
bL
DISPLAY BACKLIGHTING
By pressing
ZERO
it’s possible to programme the functioning mode of the backlighting.
Press
TARE
to change the options:
EL oFF
backlighting always off
EL on
backlighting always on.
EL AU
backlighting on only when the weight is greater than 20d or when a key is pressed. When a key is pressed,
if the weight is less than 20d, the backlighting remains active for 5 seconds.
Confirm with
ZERO
or press
PRINT/M+
to exit without confirming.
bEEP
ACOUSTIC ALARM
By pressing
ZERO
it’s possible to decide whether the acoustic alarm should enable when the “OK” symbol is on or
when the “HI” or “LO” symbols are on.
Press
TARE
to change the options:
bP 1
acoustic alarm disabled
bP 2
active acoustic signal with active “OK” symbol.
bP 3
active acoustic signal with active “HI” or “LO” symbol.
Confirm with
ZERO
or press
PRINT/M+
to exit without confirming.