Maintenance
21
Calibration
Calibration of the Bird APM-16 will be required after replacing the meter or the
instrumentation module, and also when the calibration certification expires.
Note:
Calibrating the unit, or disturbing the calibration label,
within the one-year warranty period will void the warranty.
Calibration Environment
For best results, make sure that the workspace meets the following
requirements:
The workspace should be free from electrical noise and radiated
signals.
The workspace and equipment should be at a uniform and stabilized
temperature, between 20 and 25°C (68 to 77°F).
The relative humidity should be less than 50% and noncondensing.
Equipment Required
To calibrate the Bird APM-16, you will need the following equipment:
Standard digital multimeter with DC voltmeter resolution of 0.01 V or
better (Fluke Model 87)
Bird 4401A200 plug-in calibration element
Regulated DC power supply, supplying 2 ± 0.1 V
50 ohm coaxial cables (RG-58-U), no more than 3 feet (1 m) long
BNC “T” adapter
Figure
11
Calibration Setup