Rectangular Pulses
When measuring rectangular pulse power, note the following signal
requirements:
•
Duty Cycle: 2%, minimum
•
Rep. Rate: 100 pps, minimum
•
Pulse Width: 200 µs, minimum
Peak Calibration
If recalibration is required, refer to figure 15 and these instructions:
1. Unscrew the four 8-32 flat head screws securing the back cover.
2. Grasp the back cover by the side tabs behind the line connectors
and pull straight back to remove it.
3. Insert the Bird 43P between a reliable CW signal source and a 50
ohm load.
4. Insert an element with the appropriate power and frequency
range into the Bird 43P. Turn it to read forward power.
5. Set the wattmeter to CW mode (LED off).
6. Turn on the signal source and adjust the source power to give a
stable reading in the upper half of the wattmeter’s scale.
7. Switch the wattmeter to PEAK mode (LED on).
8. Adjust the potentiometer (R9) until the PEAK reading is the same
as the CW reading.
9. Replace the back cover.
Содержание Thruline 43
Страница 2: ...THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 14: ......
Страница 17: ...Bird Model 43 Wattmeter 3 Figure 2 Bird 43 Thruline Wattmeter Outline Drawing ...
Страница 18: ...Bird Model 43 Wattmeter 4 ...
Страница 23: ...Bird Model 43 Wattmeter 9 Figure 4 Percent Reflected Power vs VSWR 1 0 1 3 ...
Страница 24: ...Bird Model 43 Wattmeter 10 Figure 5 Percent Reflected Power vs VSWR 1 0 8 0 ...
Страница 30: ...Bird Model 43 Wattmeter 16 ...
Страница 34: ...Bird Model 43 Wattmeter 20 ...
Страница 37: ...Bird Model 43 Wattmeter 23 Figure 9 VSWR Conversion Graph Reflected Power 0 2 20 0 ...
Страница 38: ...Bird Model 43 Wattmeter 24 Figure 10 VSWR Conversion Graph Reflected Power 0 01 1 00 ...
Страница 40: ...Bird Model 43 Wattmeter 26 Figure 11 Cable Length Wavelength Matching ...