FlightHawk RF Test Set
A-6
3) CALIBRATE: For best results, you can set the frequency and
calibrate the Bird Site Analyzer immediately before taking
measurements.
NOTE: Use a precision open, short, load (OSL)
calibration standard or “Calibration Combo" for accurate
measurement results.
NOTE: When using an extension cable, a phase stable
cable is needed to ensure performance and accuracy.
Connect the RF test cable and the adapter (extension)
and then calibrate to the end of this extension. If the
cable or adapters are removed, you must calibrate
again.
NOTE: Changing frequency settings will automatically
turn calibration off. Always set the frequency before
calibrating the unit.
a) Push the Calibrate icon.
b) Connect the Calibration Combo “Open” connector to
the end of the extension cable’s bench connector
adapter.
c) Push the Open softkey.
d) Wait for the unit to sweep and a Green Check to
display.
e) Connect the Calibration Combo “Short” connector to
the end of the extension cable’s bench connector
adapter.
f) Wait for the unit to sweep and a Green Check to
display.