10

Chapter 4

Theory of Operation

Measurement Process

The 7027 measures the average power of up to four states within a pulse 
period, see 

Figure

  8

. Each state is defined with begin and end times. The begin 

and end times are referenced to the internal reference signal or may be 
referenced to an external sync pulse. 

Figure

 

8  

 

State Average Measurements

The average pulse power measurement interval for each state begins at the 
user specified begin time and ends at the user specified end time. There may be 
four different states specified, each with a begin and end time. 

Содержание 7027 Series

Страница 1: ...right 2018 by Bird Technologies Inc Instruction Book Part Number 920 7027S Rev A Thruline is a registered trademark of Bird Electronic Corporation Time Domain Pulse Power Sensor 7027 Series Operation Manual ...

Страница 2: ... Do not attempt to remove the RF transmission line while RF power is present WARNING Do Not Service Or Adjust Alone Under no circumstances should any person reach into an enclosure for the purpose of service or adjustment of equipment except in the presence of someone who is capable of rendering aid WARNING Safety Earth Ground An uniterruptible earth safety ground must be supplied from the main po...

Страница 3: ...s on the equipment indicating there is important information in the instruction manual regarding that particular area Note Calls attention to supplemental information Warning Statements The following safety warnings appear in the text where there is danger to operating and maintenance personnel and are repeated here for emphasis WARNING Never attempt to connect or disconnect RF equipment from the ...

Страница 4: ...N DE SEGURIDAD DEL INSTRUMENTO BENUTZUNG WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM HERSTELLER BESCHRIEBEN KANN DIE GERÄTESICHERHEIT BEEINTRÄCHTIGT WERDEN UTILISATION TOUTE UTILISATION DE CET INSTRUMENT QUI N EST PAS EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT ENDOMMAGER LE DISPOSITIF DE PROTECTION DE L INSTRUMENT IMPIEGO QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN MODO DIVERSO DA COME SP...

Страница 5: ...GERÄTES GELTEN NUR FÜR GESCHULTES FACHPERSONAL ZUR VERMEIDUNG GEFÄHRLICHE ELEKTRISCHE SCHOCKS SIND WARTUNGSARBEITEN AUSSCHLIEßLICH VON QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN ENTRENTIEN L EMPLOI DES INSTRUCTIONS D ENTRETIEN DOIT ÊTRE RÉSERVÉ AU PERSONNEL FORMÉ AUX OPÉRATIONS D ENTRETIEN POUR PRÉVENIR UN CHOC ÉLECTRIQUE DANGEREUX NE PAS EFFECTUER D ENTRETIEN SI L ON N A PAS ÉTÉ QUALIFIÉ POUR C...

Страница 6: ...our Solon Ohio factory This manual may be periodically updated When inquiring about updates to this manual refer to the part number and revision on the title page Literature Contents Chapter Layout Introduction Describes the features of the Bird 7027 Series Power Sensor Installation Describes how to connect Bird 7027 Series Power Sensor to the user s system and software requirements Operation Desc...

Страница 7: ...l Description 1 Measurement 1 Sync Pulse 2 Data Communication Options 2 Status Indicator 2 Chapter 2 USB Model Pulse Sensor 3 Monitor Control 3 Measurement 3 Power Requirements 4 Software 4 Interlock Switch 4 Interlock wiring 5 Installing the 7027 USB Model Pulse Power Sensor 5 Install optional interlock bracket 6 Chapter 3 Ethernet Model Pulse Sensor 7 Monitor Control 7 Measurement 7 Power Requir...

Страница 8: ...i Installing the 7027 Ethernet Model Pulse Power Sensor 9 Chapter 4 Theory of Operation 10 Measurement Process 10 Chapter 5 Maintenance 11 Customer Service 11 Shipment 11 Specifications 12 Model Identification 14 Limited Warranty 15 ...

Страница 9: ...nt The 7027 is capable of measuring multi level pulsed waveforms It measures the average power gated power rise and fall times of up to four levels states within a pulse period See Figure 1 Figure 1 Multi level Pulsed Waveform Measurement See Chapter 4 Theory of Operation on page 10 for a complete explanation of the measurement process The input and output connectors of the Bird 7027 Series Power ...

Страница 10: ...d Commands for Programmable Instrumentation SCPI over USB See USB Model Pulse Sensor on page 3 Ethernet The Ethernet communications model is controlled using Standard Commands for Programmable Instrumentation SCPI over Ethernet See Ethernet Model Pulse Sensor on page 7 Refer to the 7027 Pulse Sensor Programming manual for details regarding SCPI commands Status Indicator A status indicator is provi...

Страница 11: ...mable Instrumentation SCPI over USB Refer to the 7027 Pulse Sensor Programming manual for details regarding SCPI commands Measurement The 7027 is a multi level pulse power sensor It measures the average power of up to four levels within a pulse period Each state is defined with begin and end times within the VPM3 user interface State Average Measurement The begin and end times are referenced to th...

Страница 12: ...contains a normally open interlock switch for interlock control of the connected RF source An optional bracket is available that when installed closes the interlock circuit When installed the interlock bracket ensures the RF cables cannot be disconnected from the sensor s RF input and output connectors see Figure 3 When the RF source is connected to the interlock circuit via the sensor s DB9 conne...

Страница 13: ...ance Note Connect the 7027 Power Sensor to the RF line so that the arrow on the sensor points towards the load 2 Connect the USB Type A flat connector to the PC 3 Connect the USB Type B square connector to the power sensor 4 Open Devices and Printers dialog box Control Panel Hardware and Sound Devices and Printers 5 Verify the Power Sensor s device drivers are installed correctly Note Refer to the...

Страница 14: ...forming this procedure 1 Connect the RF source interlock cable to the DB9 connector on the 7027 sensor See Interlock wiring on page 5 for connector pinout 2 Position the interlock bracket on the back of the sensor so the input and out put connectors are covered See Figure 3 on page 4 3 Tighten the two thumb screws to secure the bracket to the sensor ...

Страница 15: ... 7027 Pulse Sensor Programming manual for details regarding SCPI commands Measurement The 7027 is a multi level pulse power sensor It measures the average power of up to four levels within a pulse period Each state is defined with begin and end times The begin and end times are referenced to the internal reference signal or may be referenced to an external sync pulse See Chapter 4 Theory of Operat...

Страница 16: ...s and hold the Ethernet reset pushbutton for five seconds IP Address Configuration Note To change the sensor s IP address the PC connected to the Power Sensor must be on the same subnet 1 Connect an Ethernet cable to the PC 2 Connect the remaining end of the Ethernet cable to the power sensor 3 Open a web browser on the PC 4 Type the Sensor s default IP Address into the browser s address box then ...

Страница 17: ...ote Connect the 7027 Power Sensor to the RF line so that the arrow on the sensor points towards the load 2 Connect an Ethernet cable to the PC 3 Connect the remaining end of the Ethernet cable to the power sensor 4 Connect the 5 VDC Power supply output to the power input connector on the Power Sensor 5 If an external sync pulse signal will be used connect the BNC plug on the sync pulse input cable...

Страница 18: ...d end times The begin and end times are referenced to the internal reference signal or may be referenced to an external sync pulse Figure 8 State Average Measurements The average pulse power measurement interval for each state begins at the user specified begin time and ends at the user specified end time There may be four different states specified each with a begin and end time ...

Страница 19: ...any reason request an Return Material Authorization RMA through the Bird Technologies website All instruments returned must be shipped prepaid and to the attention of the RMA number Bird Service Center 30303 Aurora Road Cleveland Solon Ohio 44139 2794 Fax 440 248 5426 E mail bsc birdrf com For the location of the Sales Office nearest you visit our Web site at http www birdrf com Shipment Should yo...

Страница 20: ...andwidth VSWR Range 1 0 1 to 2 0 1 Gated Average Power Accuracy 1 at calibration frequencies 2 at all other frequency within the sensor bandwidth Insertion loss Max 0 05 dB Insertion VSWR Max 1 05 RF Connectors QC Overall power will be limited by connector selection See Model Identification on page 14 for QC connectors supplied by model number Directivity Min 28 dB Sync Input Input port impedance ...

Страница 21: ...nectors Weight Max Less than 3 lbs Mechanical Shock and Vibration Designed to meet MIL PRF 28800F class 3 Compliance Designed to Comply with EMC Directive 2014 30 EU European Standard EN 61326 Electrical Equipment for measurement control and laboratory use EMC Requirements Test Spec for radiated immunity EN 61000 4 3 Testing and measurement techniques 10V meter RoHS Compliant CE Mark Compliant Com...

Страница 22: ...Sensor 7027 14 Model Identification 0 12 3 45 12 5 12 12 4 5 5 4 3 4 6 3 7 8 8 3 7 4 96 8 7 96 8 6 1 4 6 1 4 9 0 3 4 4 9 0 3 4 8 4 8 6 4 8 2 3 4 8 4 8 4 8 2 4 4 8 4 4 8 A 4 8 A 4 6 8 A 2 4 B8 4 B8 6 4 B8 3 4 B8 4 4 4 4 ...

Страница 23: ...ry such claim for breach of these warranties shall be deemed to be waived by Buyer unless made in writing within ten 10 days from the date of discovery of the defect The above warranties shall not extend to any products or parts thereof which have been subjected to any misuse or neglect damaged by accident rendered defective by reason of improper installation or by the performance of repairs or al...

Отзывы: