18
Heart wire with break-off needle or with flexible end
(max. 2.3 mm in diameter)
Implanted lead with IS-1 connector
Note:
For dual-chamber cables (PK-141, PK-175, PK-67-S/L),
Reocor S uses only the ventricular channel!
Polarity
Reocor S principally paces in bipolar mode, but it can be used
with bipolar or unipolar temporary pacing leads.
If unipolar leads are used, two leads must be connected.
Separating connections
Disconnect patient cables from the temporary pacing leads of
the patient or disengage the direct connection.
Separating Redel plug
•
Retract the retaining ring at the Redel plug and pull the
Redel plug off the Redel port.
Patient-side
connection
BIOTRONIK cable
Device-side
connection
Reocor S connection
Recommended connections
Screw terminals
PK-83B
Redel plug
Redel adapter
Screw terminals
PK-83
2-mm plug
2-mm connector port
Possible connections
Screw terminals
PK-175
Redel plug
Redel adapter
Alligator clips
PK-141
Redel plug
Redel adapter
Alligator clips
PK-67-S/L with PA-4
Redel plug
Redel adapter
Alligator clips
PK-67-S/L with PK-155
Redel plug
Redel adapter
Alligator clips
PK-82
2-mm plug
2-mm connector port
Connection
patient-side
BIOTRONIK cable
Connection
device-side
Reocor S
connection
Recommended connections
IS-1 connector
port
PK-67-S/L with PA-2
Redel plug
Redel adapter
Possible connections
Alligator clips
PK-141
Redel plug
Redel adapter
Alligator clips
PK-67-S/L with PA-4
Redel plug
Redel adapter
Alligator clips
PK-67-S/L with PK-155
Redel plug
Redel adapter
Alligator clips
PK-82
2-mm plug
2-mm connector port
394270--J_GA_Reocor_S_mul.book Page 18 Monday, January 26, 2015 4:39 PM