20
Philos DR-T Technical Manual
Sensitivity
A
0.1...(0.1)...
1.0
...(0.1)…1.5…(0.5)...7.5 mV
V
0.5...(0.5)...
2.5
...(0.5)...7.5 mV
Refractory period
5
A
200...(25)…
425
...(25)...775 ms
V
170, 195, 220,
250
; 300; 350; 400 ms
PVARP (minimum)
Off,
dependent on TARP and AV Delay settings
ARP Extension
0
...(50)...350 ms
Lead Check
Off
; On
Automatic Mode Conversion
Off
; On (in modes DDD(R), DDT(R)/A, DDT(R)/V, and VDD(R))
Mode Switch (X out of Y)
Off
; On (in modes DDD(R), DDT(R)/A, DDT(R)/V and VDD(R))
X
= 3...(1)...8
Z
= 3...(1)...8
Lead Polarity
Pace: A
unipolar
; bipolar
V
unipolar
; bipolar
Sense: A
unipolar
; bipolar
V
unipolar
; bipolar
Far-Field Blanking
50...(25)...200 ms
5
In the DDIR, VVIR and VOOR modes, lower maximum sensor rates result than indicated
here (partly depending on the selected AV interval). The correct values are indicated by
the programmer.